17
3.2.3 Installation
3.2.3.1 Initial Check Out
Unpack the Control Unit and Sensor and inspect for visible
damage. If no damage is observed then proceed to check
out the system as follows:
1. The Humidity Sensor is already connected to the Control
Unit when shipped.
2. Connect the line cord into the appropriate power outlet
(MAINS). Be sure the POWER switch is in the OFF
position and the DEHUMIDIFIER switch is in the OFF
(out) position. Plug the Dehumidification System into the
AC outlet.
NOTE:
All ETS operating systems utilize the standard 3-prong
North American plug. A 230V/50Hz system will be
marked as 230V, but will still have the same 3-prong
plug
.
3. Turn on the Controller (the POWER switch is located on
the rear panel). The LCD readout will initially read a
humidity level much lower than ambient and the Green
DEHUMIDIFIER indicator should be OFF.
4. The LCD reading will slowly start to increase until it
reaches the ambient RH level. This may take as long as
5 minutes if the ambient humidity is very high.
5. Depress the READ RH/SET RH Switch to the SET PT
position. The Display will read some level. While holding
the switch down in the SET PT position, rotate the
ADJUST Control to the desired RH level set point using a
small (1/8”/3mm) screwdriver.
6. Release the switch so that it returns to the READ RH
(out) position and note the humidity level reading on the
display. Turn on the DEHUMIDIFIER switch (Green
button out). If the set point is below ambient, the Green
DEHUMIDIFIER ON light will turn on and power will be
applied to the Dehumidifier outlet, activating the
dehumidification system. If the SET PT is above the
ambient RH reading, the Green LED indicator light will be
off.
7. The dehumidification process will start. The
Dehumidification System will remain on until the set point