ets-lindgren.com
The
relay x drive
connectors on the back panel of the remote relay will switch
the relays into position.
•
For a 24 V relay, connect pin 1 (24 V DC output) to the Common of the
relay power terminals.
•
For a 12 V relay, connect pin 2 (12 V DC output) to the Common of the
relay power terminals.
•
Connect pin 3–8 to the power terminals 1–6 of the relay.
The number of power terminals to connect depends on the type of relay you use.
For example, a 1–6 relay requires all of the pins, and a 1–3 relay requires
three pins.
Readback Function
The
relay x readback
connectors on the back panel remote relay can verify if
the external relay is set into position.
1.
Connect pin 1 (3v3 out) to the indicator common of the relay.
2.
Connect pin 2 to the first indicator of the relay.
3.
Connect pin 3 to the second indicator; continue until complete.
The number of indicator terminals to connect depends on the type of relay you
use. Pin 8 is not in use.
Set the readback function on or off in the EMCenter™ Modular RF Platform
depending on the number of relay(s) with readback function that you use.
Содержание 7001-001
Страница 6: ...vi ets lindgren com This page intentionally left blank ...
Страница 16: ...16 Specifications ets lindgren com This page intentionally left blank ...
Страница 24: ...24 EMSwitch Remote Relay ets lindgren com This page intentionally left blank ...
Страница 38: ...38 EMSwitch Command Set ets lindgren com This page intentionally left blank ...
Страница 42: ...42 Typical Data ets lindgren com ...
Страница 43: ...ets lindgren com Typical Data 43 VSWR The average power input must be reduced for load VSWR above 1 1 ...
Страница 44: ...44 Typical Data ets lindgren com This page intentionally left blank ...