ENGLISH
11
INSTALLATION
CE conductivity measurement range and K
factor (cell constant)
▪ K1 Range up to 20.00 mS
Hysteresis / PWM start-up point = 500 μS
Resolution 10 μS
▪ K5 Range up to 2.,000 μS / Hysteresis / PWM start-up point = 50 μS
Resolution 1 μS
▪ K5 Range up to 200 μS / Hysteresis / PWM start-up point = 0.50 μS
Resolution 0.1 μS
Temperature settings:
Manual or automatic offset (auto with temperature probe PT100)
▪ Resolution 0.1% °C
▪ Precision: ± 0.5% °C
Temperature probe range:
– 20.... 100°C
Power supply / Consumption:
Universal power supply 100
�
250VAC / 5W at 240VAC
Microprocessor technology:
SMD components with a 6-key digital control keypad
Linearity, Stability, Reproducibility:
�
0.5% in standard conditions
Display:
Backlit 126x64 display; Visible display area 70x37 mm
Delay on Setpoint:
Relay activation delay, programmable for each setpoint (999 sec.)
Delay on start-up:
Delay in relay when the unit is switched on, programmable
Consumption / Rated Current:
230Vca 5W = 25mA ▪ 24Vca-cc=5W = 230mA ▪ 12Vcc 5W = 460mA
Internal electrical protection:
Power supply unit assures electrical protection (instead of fuse)
Level / Relay remote control:
Chemical additive level (level probe not included) output v5VDC
RELAY 1 output:
SV (solenoid valve) PURGE ON-OFF / voltage-free, relay 5Amax 230VAC
RELAY 2 output:
INHIBITOR ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
RELAY 3 output:
BIOCIDE1 ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
RELAY 5 output:
BIOCIDE2 ON-OFF metering pump/ voltage-free, relay 5Amax 230VAC
AUX Output:
ALARM voltage-free contact, relay 5A max 230Vac
FLOW sensor:
Blocks outlet operations if there is no flow in the probe socket.
Outputs: Output 0/4...20 mA1 EC:
Adjustable (500
�
maximum input impedance), with galvanic separation.
Connected to Conductivity measurement settings.
Output 0/4...20 mA2 EC:
Adjustable (500
�
maximum input impedance), with galvanic separation.
Connected to Conductivity measurement settings.
Load:
Resistive load 5A at 230VAC / Inductive load 0.5A at 230VAC
Relay insulation voltage:
> 3000VAC
Contact relay duration:
�
5x10
4
operations (5A at 230VAC)
Noise level:
Irrelevant
Operating temperature:
ideal temperature 5°C-40°C, resistance up to 0°C-45°C
Environmental conditions:
Possibly dry environment, altitude up to 2000m, Relative humidity 80% for temperature up to
31°C linearly decreasing to 50% of relative humidity at
40°C. Pollution degree 2.
Transport / storage conditions:
– 5
�
60°C in a dry environment
Install
the instrument in a dry place, away from heat sources at
a maximum room temperature of 40°C.
Comply with standards in force in the different countries
regarding electrical installation (Fig. 2).
If the power cord does not
have a plug, the device must be connected to the mains by
means of an omnipolar disconnecting switch with at least 3
mm between the contacts. All the power supply circuits must
be interrupted before accessing the connection devices
.
Fig. 1
–
Electrical connection
TECHNICAL SPECIFICATIONS OF THE INSTRUMENT