8.5 HD-PH/P
(
COMMISSIONING THE PUMP
)
a. Install the metering pump as described in Chapter 6.0 “P
UMP INSTALLATION
”.
b. Place the electrode in the electrode holder.
c. Connect the electrode to the metering pump by means of the male BNC connec-
tor (12), rotating it through 90°C.
d. Position the "FUNCTION SELECTOR" (14) at pH set and then use the screwdriv-
er supplied as part of the kit to adjust the "pH set" trimmer to the desired pH
value.
e. Position the function selector (14) at “pH met”. Use the "ACID/ALK" switch (15)
to select the required type of intervention (acid or alkaline). Example: If the
intervention point is set at pH 9 and an acidification is required, set the switch
to ACID. The pump will then start dosing the additive every time the pH rises
above 9. If it is desired to alkalinize the system, set the switch to ALK, in which
case the pump will come into operation every time the pH drops below 9.
f. Since the temperature of the liquid exerts an influence on the measured pH
value, set the " temperature compensation trimmer" (5) to the temperature exist-
ing in the system to obtain an automatic correction of the pH value ascertained
40
ENG
1
Suction hose connector nipple
2
Delivery hose connector nipple
3
Stroke Led,red
4
Power Led, green
5
Temperature compensation knob
7
Warning signaling Led, red
8
Level signaling Led, yellow
10
Level probe connector
11
Control panel
12
Electrode cable connector
13
Set-point selector “pH set”
14
Functions selector
“pH met” display the measured
pH value
“pH set” display set-point value
15
Acid/Alk switch
16
Calibration trimmer “pH/mV”
17
Calibration trimmer “
∆
0”
18
Display
19
Supply cable
20
Auxiliary output connector
fig.12
Содержание HD-PH
Страница 2: ......
Страница 4: ......
Страница 85: ...Allegato Disegni Drawings Annexe des Dessins Dibujos Anexos 85...
Страница 86: ...86...
Страница 90: ...90 6401 6101 6501 6301 2401 1901 1801 Exx ELETTROMAGNETE COMPLETO COMPLETE ELECTROMAGNET...
Страница 91: ...NOTE...
Страница 92: ...NOTE...
Страница 93: ...NOTE...
Страница 94: ...NOTE...
Страница 95: ......
Страница 96: ...COD 4063 052 00...