8.3 HD-RX
(
COMMISSIONING THE PUMP
)
1
Suction hose connector nipple
11
Control panel
2
Delivery hose connector nipple
12
Electrode cable connector
3
Stroke Led, red
13
Set-point selector “mV set”
4
Power Led, green
15
Red/oxid switch
5
Injection frequency regulation knob
16
Calibration trimmer “/mV”
6
Front cover fixing screw
17
Calibration trimmer “
∆
0”
9
Fuse
18
Display
10
Supply cable
a. Install the metering pump as described in Chapter 6.0 "P
UMP INSTALLATION
".
b. Place the electrode in the electrode holder.
c. Connect the electrode to the metering pump by means of the male BNC connec-
tor (12), rotating it through 90°C.
d. Position the "mV set" knob (13) to the desired Rx value, bearing in mind that the
scale dimensions are such as to give rise to possible slight discordances
between the intervention point and the reading on the display (18). For exam-
ple, if the "mV set" knob is set to 700 mV and one then notes that pump has not
yet come into operation 720 mV, it would be desirable to carry out a check and,
if necessary, correct the set point by adjusting the "mV set" knob.
e. Use the "RED/OXID" switch (15) to select the required type of intervention
(reduction or oxidation). In this case the following must be taken into account,
an oxidant product increase the value and a reductant product decreas the
value. Example: If the intervention point is set at 700 mV and an oxidation is
required, set the switch to OXID. The pump will then start dosing the additive
every time the solution potential drops below 700 mV. If it is desired to perform
a reducing action, set the switch to RED, in which case the pump will come into
operation every time the solution potential rises above 700 mV. Bear in mind
that Redox measurements are affected by pH variations; if this factor is subject
to considerable oscillations, we would advise you either to supplement the HD-
38
ENG
fig.10
Содержание HD-PH
Страница 2: ......
Страница 4: ......
Страница 85: ...Allegato Disegni Drawings Annexe des Dessins Dibujos Anexos 85...
Страница 86: ...86...
Страница 90: ...90 6401 6101 6501 6301 2401 1901 1801 Exx ELETTROMAGNETE COMPLETO COMPLETE ELECTROMAGNET...
Страница 91: ...NOTE...
Страница 92: ...NOTE...
Страница 93: ...NOTE...
Страница 94: ...NOTE...
Страница 95: ......
Страница 96: ...COD 4063 052 00...