ETAS
Handling of Problems
ES930.1
-
User Guide
78
Data losses (packet
losses) occur during the
transfer.
Are you using WLAN in
your measuring setup?
WLAN is not allowed
within this ETAS net
-
work. Use only ETAS
cables for the cabling of
your measuring setup
(ETAS modules and their
cable connection to the
PC).
Are you using the correct
network card type in your
laptop?
Check whether you are
using a PCMCIA network
card in your laptop. PCM
-
CIA cards with 8-bit or
16-bit data bus are not
suitable. Use only PCM
-
CIA cards with 32-bit
data bus, mini-PCI or
ExpressCards.
Are you operating the
ES930.1 on a PC with
multi-core processor?
Install a current driver for
the network card that
supports the NDIS proto
-
col.
The LED lights red.
Did you just perform an
update?
For users of the INCA/
INTECRIO/ ASCET-RP/
Configuration tool:
Switch the module off
and on again.
For users of the Config
tool:
Switch the module off
and on again.
Load the measuring con
-
figuration again.
If the LED is still lit, return
the module to ETAS for
repair.
The firmware of one or
several modules cannot
be updated.
Is the module to be
updated in a module
chain?
Update the firmware of
these ETAS daisy chain
modules separately.
1)
: The manufacturers of the PC cards refer to this function in different ways.
Example: "Link down Power saving"
Problem
Diagnostics questions
Possible solutions