BM04101 Molift Quick Raiser English - Rev F/ 09/2012
Page 10 of 20
Safety devices
Molift Quick Raiser is equipped with several safety
devices, which are intended to prevent damage to
personnel and equipment in case of incorrect use.
• The lifter has an overload sensor preventing
the lifter to be operated with more than SWL.
• The lifting arm is hinged to prevent the arm
from squeezing the patient if it is lowered too
much.
• The lifting arm has rope locks on the lifting
arms, designed to lock the ropes on the lifting
strap when they are loaded.
• When activating the emergency stop button,
the power supply will be disconnected, and the
lifter will stop.
• If an error occurs, it is possible to lower the lift-
ing arm manual or electrical.
• The lifter has a service lamp which gives a sig-
nal when the lifter needs service.
Emergency stop and emergency lowering
Contact a service partner if the cause of the
stop or malfunction is unknown or if a fault is
discovered. The lifter is to be taken out of serv-
ice until the fault is repaired
Emergency stop
The emergency stop button is placed on the chas-
sis. The emergency stop will cut the power sup-
ply to the motor when it is activated. Activate by
pushing the emergency stop button in, and re-
set by turning the button clockwise until it pops
up. Electrical emergency lowering does not work
when emergency stop is activated.
Emergency lowering
It is possible to perform a emergency lowering if
the lifter should malfunction. Molift Quick Raiser
has both manual and electrical emergency lower-
ing.
The manual lowering is done by turning the worm
screw using an Allen key (this can be found on the
top of the column). Puncture the label and use the
small Allen key in the hole under the yellow label,
turn the worm screw until the patient is lowered
enough.
Electrical emergency lowering can be found on
the chassis. Push the emergency lowering button
and lower the patient. Electrical emergency low-
ering does not work when the emergency stop is
activated.
Replace labels if they are punctured to seal
the lifter again.
Emergency
stop
Electrical emer-
gency lowering
Manual
emergency
lowering
Allen key for
manual emer-
gency lowering
Allen key for
column fixing
bolts