Transport and operating conditions
Transport and Storage
For long time storage it is recommended that the
emergency stop button is activated (pushed in).
The lifter can be stored and transported under
temperatures between -25 - 70 °C.
Operating
The lifter is designed for use at standard room
temperatures (+5 to +40°C).
Air Pressure: 70 - 106 kPa
Relative Humidity: 15 - 93 %
Storage or transport at other temperatures leave
the lifter in a room with a suitable temperature
until it reaches a safe operating temperature.
Medical electrical equipment
requires special precautions regard-
ing electromagnetic compatibility
(EMC). Portable or mobile radio
communication equipment may
affect the medical electrical equip-
ment, and should be kept minimum
25 cm (10 inches) from the lifters
electronics.
Battery
Molift Quick Raiser 205 is available with a 24V SLA
battery or a 25.6V LiFePO4 battery - both with an
integrated charger.
The SLA battery has a life expectancy of approx.
500 charge cycles.
The LiFePO4 battery has a life expectancy of
approx. 2000 charge cycles.
Recharge battery upon receipt, and
then every six months, to avoid
battery failure.
Batteries must be disposed of as
special waste according to local
rules and regulations. Do not
dispose in general waste
Battery and Service indicator
Wrench symbol /
Service indicator
Battery symbol /
Battery indicator
The electrical system has a power-save function
which will turn off the electrical system five
minutes after last operation. All indicators will turn
off.The system is activated when pushing one of
the operating buttons
Battery indicator on control panel
When the multi level battery indicator is yellow
the lifter will have sufficient power available for
one full lifting cycle with max load. When battery
is flashing yellow it is only possible to lower the
lifting arm.
Battery indicator
Battery level
Green
80% or more
Green
80 - 60%
Green
60 - 40%
Green
40 - 20%
Yellow or
yellow
blinking
20% or less
(recharge lifter)
Electronics
14
Molift QuickRaiser 205 / www.etac.com
Содержание Molift Quick Raiser 205
Страница 2: ......
Страница 3: ...Molift Quick Raiser 205 EN User manual BM13001 Rev E 2021 06 09...
Страница 22: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 23: ...Molift Quick Raiser 205 DE Bedienungsanleitung BM13002 Rev E 2021 06 09...
Страница 42: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 43: ...Molift Quick Raiser 205 NO Brukermanual BM13003 Rev E 2021 06 09...
Страница 62: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 63: ...Molift Quick Raiser 205 SE Anv ndarhandbok BM13004 Rev D 2021 06 09...
Страница 82: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 83: ...Molift Quick Raiser 205 FI K ytt ohje BM13005 Versio D 2021 06 09...
Страница 102: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 103: ...Molift Quick Raiser 205 IT Manuale dell utente BM13006 Rev D 2021 06 09...
Страница 122: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 123: ...Molift Quick Raiser 205 FR Manuel d utilisation BM13007 R v D 2021 06 09...
Страница 142: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 143: ...Molift Quick Raiser 205 ES Manual del usuario BM13008 Rev D 2021 06 09...
Страница 162: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 163: ...Molift Quick Raiser 205 NL Gebruikershandleiding BM13009 Rev D 2021 06 09...
Страница 182: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...
Страница 183: ...Molift Quick Raiser 205 DA Brugervejledning BM13010 Rev D 2021 06 09...
Страница 202: ......
Страница 203: ......
Страница 204: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...