9
Molift Partner 255 / www.etac.com
Charging
Battery Charger
The battery charger can be used as a tabletop
charger or mounted to a vertical wall surface
close to a power outlet, minimum 120 cm above
the floor. The charger has to be used outside the
patient environment. The Charger must be placed
or installed in a way that makes it easy to discon-
nect mains cable plug. Take care not to damage
the cable. The charger can be connected to the
power outlet at all times.
Battery Charger 12-24V
Art. no. 1340100
Battery PowerPac 26,4 V NiMH
Art. no. 1341000
Charger Light
Description of charger light:
Charger light
Mode
Yellow
Ready for use, no bat-
tery connected
Yellow
Initialization
Red
Fast Charge
Green/Yellow
Top-off charge
Green
Trickle charge
Red/Green
Error
Integrated charger (optional)
The lifter can be charged without removing the
battery from lifter with integrated charger. Park
the lifter and connect to a wall socket to charge
the battery. Lifter can be delivered with integrated
charger or it can be delivered as an accessory and
retrofitted.
Integrated charger
Art. no. 1340101
Charger
Light
Battery
Charger
Mains cable
(disconnecting