Sense Mode
30
ET System electronic GmbH
GENERAL
INFORMATION
FOR
SENSING
The sensing line moves the measuring point of the output voltage measurement to the load. The voltage
measurement is carried out with a differential amplifier, the last resistor in the chain being "short-circuited" with a
low-resistance parallel resistor. When the sensing line is opened, the output voltage increases between 1% and 2% of
the nominal voltage, depending on the model. The displayed voltage on the display does not change. When the Sens is
connected to the load, the power resistor is mathematically transformed into the device, resulting in a residual error
of up to 0.5% depending on the load line used. More detailed information can be obtained from the manufacturer.
WARNING
INSTRUCTIONS
FOR
USING
RELAYS
TO
THE
LOAD
DRAFT
In an application where the load is for example should be dropped with a relay and the Sens is used at the same time,
care must be taken to ensure that the load current can not flow through the sensing lines as this can lead to the
destruction of the sensor. This is shown schematically in the following figure.
The circuit breaker (S_power) must be closed when starting before the Sens (S_sens). When switching off, the Sens
must first be opened and then the circuit breaker can be opened. Otherwise, a current flow may occur across the
sensing line, and this may be particularly critical when disconnecting.
Содержание LAB-HP/E 10KW
Страница 2: ...Software version V15 November 2016...
Страница 5: ......
Страница 6: ......
Страница 10: ...4 ET System electronic GmbH IMPORTANT SAFETY INSTRUCTIONS...
Страница 12: ...Important Safety Instructions 6 ET System electronic GmbH...
Страница 40: ...Notes 34 ET System electronic GmbH NOTES...
Страница 41: ...Notes ET System electronic GmbH 35...