Technical Specifications
ET System electronic GmbH
3
O
UTPUT SPECIFICATION
:
Static Regulation
Line Regulation Voltage
Line Regulating Current
Load Regulation Voltage
Load Regulation Current
Dynamic Response Time
@Load Step 10 - 90%
Output Voltage [V]
15
20
25
35
40
45
70
80
100
150
300
600
800
1000 1200 1500
HP/E 5KW Unit [A]
x
250 200 150
125
120
75
65
50
35
17
8,5
6,5
5
4,2
3,3
HP/E 10KW Unit [A]
x
500
x
350
250
250
175
130
100
75
40
17
13
10
8
6,6
HP/E 15KW Unit [A]
x
750
x
x
375
x
x
195
150
100
50
25
19
15
12
9,9
Voltage ripple (p-p)
20MHz
40
12
80
12
80
12
80
12
140
12
140
12
140
12
140
12
140
12
900
12
900
12
900
12
1000
12
1200 2500 2500
Voltage ripple (p-p)
300KHz
15
12
35
12
35
12
35
12
60
12
60
12
60
12
60
12
60
12
400
12
400
12
400
12
700
12
800 1500 1500
Voltage ripple (rms
)
5
mV
20MHz
15
35
35
35
60
60
60
60
60
400
400
400
400
400
400
500
Voltage ripple (rms
)
5
mV
300KHz
10
12
25
12
25
12
25
12
40
12
40
12
40
12
40
12
40
12
300
12
300
12
300
12
300
12
300
300
400
Current ripple (p-p)
Current ripple (rms)
Primary / Sekundary
DC - Output/ Earth
Primary / Earth
Rise time full Load
13
6 ms
6 ms
6 ms
6ms
12 ms
12 ms
12 ms
20 ms
20 ms
20 ms
20 ms
20 ms
40 ms
40 ms 40 ms 6 ms
Rise time no Load
13
5 ms
5 ms
5 ms
5 ms
10 ms
10 ms
10 ms
10 ms
10 ms
10 ms
10 ms
10 ms
10 ms
20 ms 20 ms 5 ms
Fall time full Load
13
15 ms 15 ms 15 ms 15 ms 20 ms
20 ms
20 ms
20 ms
20 ms
40 ms
40 ms
50 ms
60 ms
80 ms 100 ms 25 ms
Fall time no Load
Rel. Accuracy ±[%]
7
Voltage [V] 0,25
0,038 0,050 0,063 0,088
0,100
0,113
0,175
0,200
0,250
0,375
0,750
1,500
2,000
2,500
3,000 3,750
Current [A] 0,4
1,000 1,000 0,800 0,580
0,500
0,480
0,288
0,260
0,200
0,140
0,068
0,034
0,026
0,020
0,017 0,013
Voltage [V] 0,5
0,075 0,100 0,125 0,175
0,200
0,225
0,350
0,400
0,500
0,750
1,500
3,000
4,000
5,000
6,000 7,500
Max. Sens Voltage over
nominal Voltage
7
Max. Sens Voltage inside the
normal Voltage range
7
Resolution Voltage
Display
Voltage setting resolution
Resolution Current
Display
Current setting resolution
OVP
UI-Modus
100V - 999,9V
000.0
15V - 99,99V
00.00
10A - 99,99A
00.00
100A - 999,9A
000.0
1.000A - 9.999A
0000
Over Voltage Protection: is adjustable between 0% and 120% of Voltage full range
Voltage and Current operation mode: Voltage and current are setable
D
ev
ic
e Fun
ct
io
n
OTP
Over Temperature Protection: if the internal heat sink tempearatury is go above 90°C
the divice will automaticaly shut down
Reso
lu
tio
n
<0,4 % of F.S.
Rip
p
le and
No
ise
11
Accu
racy
3.000VAC
500VDC
2.000VDC
2.150VDC
Iso
latio
n
5 s unter 50VDC
Ou
tp
u
t
sp
ee
d
Relative Accuracy for Sens operation (worst case) [%]
1.000V - 1.500V
0000
0A - 9,999A
0.000
5% of F.S. (if higher voltage is needed ask manufacturer)
±1% of F.S.
OCP
Over Current Protection: is realised by the current set point.
The output current can not go over the set value
Regu
latio
n
± 0,1% of F.S.
± 0,02% F.S.
± 0,02% F.S.
± 0,05 % F.S. ± 2mV
± 0,05% F.S. ± 20mA
< 3ms
Other combinations of Voltage and Current also possible, ask the Manufakturer
Basic calculation of possible combinations are:
Pout = Vout x Iout / Pmax for one Unit 90KW / Cmax = 250A each 5KW
Outp
u
t
Cu
rren
t
<0,5 % of F.S.
Содержание LAB-HP/E 10KW
Страница 2: ...Software version V15 November 2016...
Страница 5: ......
Страница 6: ......
Страница 10: ...4 ET System electronic GmbH IMPORTANT SAFETY INSTRUCTIONS...
Страница 12: ...Important Safety Instructions 6 ET System electronic GmbH...
Страница 40: ...Notes 34 ET System electronic GmbH NOTES...
Страница 41: ...Notes ET System electronic GmbH 35...