7.2 Changing the insulin manufacturer
If you are changing the type of insulin preparation, then in addition to the steps
described in Chapter 7.1, you must select the correct holder and the correct
ESYSTA
®
Adapter (see Chapters 6.1 and 6.2). Never make changes to the
holder or ESYSTA
®
Adapter.
Change the details of the insulin in the ESYSTA
®
App and if necessary in the
ESYSTA
®
Portal.
7.3 Important safety information about changing the cartridge
After changing the insulin cartridge, you must carry out a venting
procedure. You may frequently have to carry out several venting
procedures after changing the cartridge (see Chapters 6.7 and 7.8).
You must use a new suitable ESYSTA
®
Adapter with each new insulin
cartridge!
Even if your last insulin cartridge was not completely emptied, because
your doctor changed your treatment, for example, a new ESYSTA
®
Adapter must also be used with the new insulin cartridge.