15
7-4 Error message ‘Target (Focus) Motor fault’ appears.
Main causes when error message ‘Target (Focus) Motor fault’ appears:
・
Failure of control board
・
Failure of optical unit
Is ‘Focus (Target) Motor
fault’ still displayed?
‘Target (Focus) Motor fault’ is displayed.
OK
Replace the optical unit ASSY.
Replace the control board.
Replace the optical unit
ASSY and control board.
Replace the harness connected to CN703
with the one connected to CN704.
Are both ‘Target Motor fault’ and
‘Focus Motor fault’ displayed?
YES
YES
NO
NO