1
Reminder - cylinder axis search
The search for the cylinder axis consists of comparing two positions:
1. The negative axis of the corrective cylinder
2. The cylinder axis of the patient correction
If the axis of the correction is correct, the patient does not perceive any difference between the two
positions.
However, if the patient perceives a difference between the two positions, the correction axis must be
adjusted by 5°
(*)
in the direction of the negative axis of the preferred cross cylinder. The operation
must be repeated until the patient no longer perceives a difference between the two positions or indicates
a return to the previous axis position.
Reminder Cylinder power search
The search for the cylinder power consists of positioning the meridians of the cross cylinder according to
the direction of the axis of the correction and comparing the two positions of the cross cylinder.
If the power of the cylinder is correct, the patient does not perceive a difference.
However, if the patient perceives a difference it is necessary to modify the power of the cylinder. If the
patient prefers:
The position of the cross cylinder with the negative axis aligned with that of the correction: it is
necessary to
increase
the negative cylinder value of the correction by 0.25 D
(*)
.
The position where the negative axis of the cylinder is perpendicular to the axis of the correction
(corresponds to the positive cylinder axis aligned with that of the correction): it is necessary to
reduce
the cylinder value by 0.25 D
(*)
.
Repeat the operation until the patient no longer perceives a difference or indicates a return to the
previous position of the cross cylinder.
Note
: after a change of 0.50 D to the cylinder, do not forget to adjust the sphere power of 0.25 D in
order to maintain the constant equivalent spherical power.
Procedure - Test Performance, Step 1 Cylinder Axis Search
Press
.
This test can also be performed with a letter target
.
The reversed cross cylinder test is displayed in the display area in the bottom of the touch screen of the
console.
>
The dot test is displayed on the test presentation screen.
>
U
SER
MANUAL
> P
ERFORMANCE
OF
TESTS
DURING
A
REFRACTION
EXAMINATION
55
Vision-R 700 - Essilor automatic phoropter > V1 - 09-2020
Содержание VISION-R 700
Страница 1: ...USER MANUAL...
Страница 4: ...I INTRODUCTION...
Страница 8: ...II INSTRUMENT...
Страница 16: ...III ADJUSTMENTS BEFORE THE EXAMINATION...
Страница 24: ...IV BASIC FUNCTIONS FOR PERFORMING A REFRACTION EXAMINATION...
Страница 37: ...V PERFORMANCE OF TESTS DURING A REFRACTION EXAMINATION...
Страница 76: ...VI MEASUREMENT...
Страница 80: ...VII REFRACTION PROGRAMS...
Страница 93: ...VIII INSTRUMENT SETTINGS...
Страница 109: ...IX TROUBLESHOOTING...
Страница 111: ...X TECHNICAL DATA...
Страница 114: ...XI CAUTIONS WARNINGS...
Страница 123: ...No undesirable side effects USER MANUAL CAUTIONS WARNINGS 123 Vision R 700 Essilor automatic phoropter V1 09 2020...
Страница 124: ...XII QR CODE...
Страница 125: ...USER MANUAL QR CODE 125 Vision R 700 Essilor automatic phoropter V1 09 2020...
Страница 126: ...USER MANUAL QR CODE Vision R 700 Essilor automatic phoropter V1 09 2020 126...
Страница 127: ...USER MANUAL QR CODE 127 Vision R 700 Essilor automatic phoropter V1 09 2020...