Always use a slightly damp soft cloth (microfiber, silicone), to clean the elements of the head:
The face shields by removing them beforehand
The optics
patient side (only if a trace is identified)
practitioner side
The camera window for near-vision distance measurements
The camera windows for Vertex distance measurements
The LED panel
Do not clean the observation windows (patient side) with liquid, nor with a compress held in a clamp or a
screwdriver to prevent damage of the optical surfaces.
To clean the SCV modules (patient side observation windows):
The SCV modules need to be checked after each patient. Visually check if traces of dirt are present on the
rear window of the SCV module (patient side).
1. Take one of the cleaning swab (provided with the product).
> Change the cleaning swab for the second module.
2. Spray Isopropyl alcohol (cleaner, antiseptic and disinfectant) on the tip (white part) of the cleaning
swab.
> Do not dip or soak the cleaning swab directly in alcohol.
3. Fold the nozzle, in order to have a larger cleaning surface.
4. Apply the tip in the center of the module and clean the module with a circular motion (snail types).
> Spiral movement from the center to the outside of the module.
Do not use wipe
Do not use a tool to clean (screwdriver, pen tip)
Do not clean directly with your fingers
b. Cleaning the console
Always use a slightly damp soft cloth (microfiber, silicone), to clean the elements of the console:
The touch screen
The keyboard
Do not spray liquid on the touch screen or the keyboard of the console, regardless of the liquid, in order
not to risk damaging the electronic boards.
5. Intended use
Intended purpose: this device perform a subjective refraction.
Indications for use: the instrument allows the user to control the patient's visual acuity.
Intended patient population: any adult or child with pupillary distance from 49mm to 80mm.
Intended part of the body applied to the device: cheeks and front skin are in contact with the device. Skin
in the contact with the device must be in healthy condition without wounds, irritation or inflammation.
Intended user: this instrument is intended for a medical use and can only be used with the instructions of
visual health expert authorized by the laws in force in the country concerned, in optometrist or
ophthalmologist office.
Intended condition of use: device must be installed in a refraction environment according to
environmental conditions written in this document.
Operating principle: basic operating cycle is: patient installation / patient's eyes centering / refraction
protocol selection & launch / refraction result recovery (data export, printing or manual recording) /
removal from patient.
No contraindications.
U
SER
MANUAL
> C
AUTIONS
& W
ARNINGS
Vision-R 700 - Essilor automatic phoropter > V1 - 09-2020
122
Содержание VISION-R 700
Страница 1: ...USER MANUAL...
Страница 4: ...I INTRODUCTION...
Страница 8: ...II INSTRUMENT...
Страница 16: ...III ADJUSTMENTS BEFORE THE EXAMINATION...
Страница 24: ...IV BASIC FUNCTIONS FOR PERFORMING A REFRACTION EXAMINATION...
Страница 37: ...V PERFORMANCE OF TESTS DURING A REFRACTION EXAMINATION...
Страница 76: ...VI MEASUREMENT...
Страница 80: ...VII REFRACTION PROGRAMS...
Страница 93: ...VIII INSTRUMENT SETTINGS...
Страница 109: ...IX TROUBLESHOOTING...
Страница 111: ...X TECHNICAL DATA...
Страница 114: ...XI CAUTIONS WARNINGS...
Страница 123: ...No undesirable side effects USER MANUAL CAUTIONS WARNINGS 123 Vision R 700 Essilor automatic phoropter V1 09 2020...
Страница 124: ...XII QR CODE...
Страница 125: ...USER MANUAL QR CODE 125 Vision R 700 Essilor automatic phoropter V1 09 2020...
Страница 126: ...USER MANUAL QR CODE Vision R 700 Essilor automatic phoropter V1 09 2020 126...
Страница 127: ...USER MANUAL QR CODE 127 Vision R 700 Essilor automatic phoropter V1 09 2020...