2. E
LECTROMAGNETIC
IMMUNITY
Guidance and manufacturer's declaration – electromagnetic immunity
Retina800 is intended for use in the electromagnetic environment specified below. The customer or the end
user of the Retina800 should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±2, 4, 8, 15 kV air
±8 kV contact
±2, 4, 8, 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Transient/sequence of
rapid electrical pulses
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for I/O lines
±2 kV for power
supply lines
Not applicable
The quality of the mains voltage should
be that of a typical commercial or hospital
environment.
In case of 2 kV bursts on the power
supply line, a malfunctioning of the
selected fixation LED may occur, move
the power supply to a free socket to which
no other devices are connected.
Surge
IEC 61000-4-5
±1 kV between
phases
±2 kV between the
phase(s) and the
ground
±1 kV between
phases
±2 kV between the
phase(s) and the
ground
The quality of the network voltage should
be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
0% U
T
for 0,5 cycle
0% U
T
for 1 cycle
70% U
T
for 25
cycles
0% U
T
for 250
cycles
0% U
T
for 0,5 cycle
0% U
T
for 1 cycle
70% U
T
for 25
cycles
0% U
T
for 250
cycles
The quality of the mains voltage should
be that of a typical commercial or hospital
environment. If the Retina800 user
requires continued operation during mains
power outages, it is recommended to use
Retina800 with an uninterruptible power
supply or batteries.
High frequency
(50/60Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable, the
device does not
contain components
susceptible to
magnetic fields.
Magnetic fields should be at levels
characteristic of a typical location in a
typical commercial or hospital
environment.
Note: U
T
s the mains voltage in AC before applying the test level.
U
SER
MANUAL
> X. G
UIDANCE
AND
MANUFACTURER
'
S
DECLARATION
113
Retina800 - Fundus camera > V2 - 06-2019
Содержание Retina 800
Страница 1: ...User manual...
Страница 4: ...I INTRODUCTION...
Страница 6: ...II SAFETY...
Страница 14: ...III SUPPLY PACKAGE...
Страница 18: ...IV INSTALLATION OF THE DEVICE...
Страница 30: ...V USE OF THE DEVICE...
Страница 58: ...VI SETTINGS...
Страница 101: ...VII MAINTENANCE...
Страница 103: ...VIII TROUBLESHOOTING...
Страница 106: ...IX ASSISTANCE NOTIFICATION ALERT AND LABEL...
Страница 108: ...X GUIDANCE AND MANUFACTURER S DECLARATION...
Страница 112: ...XI QR CODE...