3.3.6 Defrost Setting
Step
Description
Button
Display
1
Press MENU button until the display shows Setting. Press SET TEMP to
enter the menu.
MENU –
SET TEMP
2
Press MENU button until the display shows Defrost.
MENU
3
Press UP/DOWN button to set the defrost timer. Press SET TEMP to save
the defrost timer setting.
UP/DOWN –
SET TEMP
4
5
Press UP/DOWN to set the auto defrosts ON/OFF. Press SET TEMP to
save setting
UP/DOWN –
SET TEMP
3.4 Calibration (Offset Range: -20
o
C – 20
o
C)
Step
Description
Button
Display
1
Press MENU button until the display shows Calibration. Press SET TEMP
button to enter the Calibration menu
MENU –
SET TEMP
2
.
3
Press UP/DOWN button to correct temperature deviation. Press SET
TEMP to save setting
UP/DOWN –
SET TEMP
General calibration instructions:
For example: set the calibration offset for correction of a temperature deviation in the load at 100
o
C
Equipment set point = 100
o
C, equipment actual display = 100
o
C, actual temperature tested= 99.6
o
C (put the
sensor of a calibrated reference instrument near to the temperature probe of the product).
Enter the Menu and find CAL (calibration) according to the steps above.
The default value is 0.0
Set calibration correction according to above steps in the table --- CAL= -0.4
o
C
Exit the menu, then the actual display will be 99.6
o
C.
Equipment will run automatically to 100
o
C to achieve the set point.
Minimum calibration value is 0.1
o
C.
Note:
if no calibration is set,
is restored as factory calibration shown on “Set Point” display.
3.5 Quick Keys (Normal Operation Mode)
UV Timer
Timer
UV1 Timer
Return to normal
Return to normal