Low Temperature BOD Incubator
–
2.1 General Requirement
2.1.1 Location Requirements
Isotherm Low Temperature BOD Incubator (IFC) must be placed on the solid and level surface. It is
important that the equipment is set up accurately horizontally and against abnormal vibration and
noise. Door may have to be adjusted. (See Chapter 4 “Maintenance and Cleaning”).
Lock the 2 front casters (by pushing down the pedals) to fix position of the equipment.
Do not place equipment outdoors, under sunlight, where be splashed onto the equipment, heavy
concentration of dust, moisture or close to inflammables!
NOTE:
When the equipment is to be moved, release all the 2 front casters (by pull up the pedals) and fully support the
equipment while moving.
2.1.2 Environmental Requirements
Indoor use
Altitudes of up to 2000 meters (6600’)
Below 80% relative humidity – non condensing
Temperature between 15
O
C - 30
O
C (59
O
F - 86
O
F)
If ambient temperature fluctuation is more than 6
o
C, the control system will be reset according to its self-
inspection. Therefore control on temperature may be influenced.
Ensure stability of ambient temperature as temperature set point is close to ambient temperature. Especially
for temperature set point = ambient temp 0~10
o
C, working in air-conditioning room is recommended;
otherwise non-erasable temperature fluctuation may be caused.
Technical specification for the equipment is specified for units with standard equipment at an ambient
tempe25
o
C and a voltage fluctuation of ±10 %.
The temperature data are determined in accordance to factory standard following DIN
12 880.