Do not spray disinfectant directly onto the CO₂, O₂ or humidity sensor.
Allow disinfectant to react as specified by manufacturer.
Wipe all surfaces using a moistened clean cloth to remove disinfectant.
Wipe all surfaces thoroughly until dry.
Note: When cleaning and disinfecting, always observe the safety instructions and hygiene
regulations.
5.
Re-assemble the unit by reversing s
Decontamination
Decontamination in CelCulture uses 90°C moist heat to kill common organisms that
contaminates workspace including all installed components. The entire program run of the
decontamination routine takes approximately 15 hours (overnight process. The chamber is clean
and dry at the end of the cycle. No further wipe down is necessary.