background image

 

CTN

 

 

L

ABORATÓRIO DE 

E

NSAIOS 

T

ECNOLÓGICOS EM 

Á

REAS 

L

IMPAS

 

 

E

STRADA 

N

ACIONAL 

10,

 

2695-066

 

B

OBADELA 

 

Edição: 01 

Revisão: 00  

Data: 2014 

 

 

HM 

Inner door opened for an extended period of 
time  

Close the door and let the unit stabilize.  

 

Access port stopper at the rear of cabinet 
was removed or not install  

 Ensure access port stopper and filter are 
installed.  

 

Air leakage through Inner door gasket  

 Check installation of the gasket ensure it is 
properly secure to the cabinet and that there 
is no gap.  

Check the joining ends of the gasket to make 
sure there is no gap. Use silicon sealant to seal 
the joins if there is a gap.  

Check gasket profile is not deformed or torn. 
Otherwise replace new door gasket.  

 

 

Temperature is above or below temperature set-point 

Possible Cause  

Corrective Actions  

 

Temperature set-point was recently lowered 
or raised. Unit is still not stabilized after set-
point was changed.

  

 

Check temperature set-point. Allow 2 hours 
for the unit to stabilize after adjustment of 
new set-point.  

If the temperature was lowered, open the 
outer and inner doors.  

Lowering the temperature without opening 
the doors could lead to condensation.  

Check if the glass door latch is properly 
closed.  

 

Inner door opened for an extended period of 
time.  

Close the door and let the unit to stabilize.  

 

 

 

 

 

Отзывы: