background image

9

Assembly and Installation

Pre-Installation Tasks

Pre-planning is essential for successful installation. Discuss requirements with your sales representative and request 
dimensional drawings prior to construction of your site.

Before installing any components, follow the safety information in the ETS-Lindgren 

Product 

Information Bulletin

 included with your shipment.

Ensure power is off and base is secured before proceeding with installation.

Positioner Installation

Proper installation of the 2301 Precision Positioner unit directly affects performance of the positioning system as well 

as the accuracy of the test results.
1.  Uncrate all parts. Check all parts for any shipping damage. Ensure a clear area is available to assemble the  
 

positioner unit safely. 

 

  

Do not discard packing material until the Turntable is fully assembled and correct operation is verified.

 

The 2301 Precision Positioner consists of:

•  Variable speed turntable
•  Lossy foam absorber to cover the positioner
•  Vertical columns with tilt positioner
•  Laser for aligning system

 

 

Customer provides RF, power, and control cabling to the top of the unit and to the EUT. 

2.  The center of rotation for the unit must intersect the line through the bore sight of the measurement antenna.  
 

Use the included laser to verify alignment.

3.  Place the unit so the positioner connections are easily accessible and located closest to available feed through  
 

panels and power supply connections. Ensure the Model 2301 is connected to a power outlet that  

 

supplies 208-230VAC 50/60Hz. Use the power cord provided with the Model 2301 to make this connection  

 

to the system. 

 

 

US shipments will include a NEMA type 6-15P electrical plug power cord. 

 

 

International shipments will include a Schuko type electrical plug power cord. 

 

 

The IEC power inlet assembly includes a filter and an illuminated power switch.

4. 

Connect the positioner to the included Ethernet-to-fiber optic converter using the included dual fiber optic cable 

 

 

with type ST connections. The converter connects to the host computer via the included 0.9 m (3.0 ft) Cat5  

 

Ethernet cable. Ensure the fiber converter TX line is connected to the positioner RX input connector, and the 

 

 

fiber converter RX line is connected to the positioner TX output connector.

Содержание ETS-LINDGREN 2301

Страница 1: ...2301 Positioner User Manual 1723531 Rev B January 2021...

Страница 2: ...does it convey any license under its patent rights nor the rights of others All trademarks are the property of their respective owners Copyright 2020 2021 by ETS Lindgren Inc All Rights Reserved No p...

Страница 3: ...ons 8 Physical Specifications 8 Electrical Specifications 8 Assembly and Installation 9 Pre Installation Tasks 9 Electrical Installation 10 Input Output Connections 11 Trigger 11 Reset 11 Fiber Optic...

Страница 4: ...uct Caution Denotes a hazard Failure to follow instructions could result in minor personal injury and or property damage Included text gives proper procedures Warning Denotes a hazard Failure to follo...

Страница 5: ...ive grounding conductor inside or outside the instrument or disconnection of the protective earth terminal could result in personal injury Warning HEAVY OBJECT Unassisted lifting can cause injury Mech...

Страница 6: ...an illuminated switch and the inlet for the customer s EUT is operated by a non illuminated switch Labels on the top surface of the motor base indicate the location of each inlet Standard Configuratio...

Страница 7: ...heck cables for wear Ensure they are clear of potential damage from moving parts Ensure bearing and pinion have adequate lubrication Replacement Parts Following are the part numbers for ordering repla...

Страница 8: ...out Base Station Mount 167 64 cm 66 0 in Height with Base Station Mount 213 11 cm 83 9 in Width 103 63 cm 40 8 in Depth 101 60 cm 40 0 in Overall Weight 90 72 kg 200 0 lb Maximum Load Capacity 24 95 k...

Страница 9: ...for aligning system Customer provides RF power and control cabling to the top of the unit and to the EUT 2 The center of rotation for the unit must intersect the line through the bore sight of the mea...

Страница 10: ...power source is appropriate for the wire gauge used and that the wire gauge is adequate for the motor load Using undersized wire gauge will result in a higher than expected voltage drop in the power...

Страница 11: ...0 Used in case the IP address of the device is unknown Fiber Optic The positioner is equipped with a fiber optic inlet and a fiber optic outlet Connect the positioner to the included Ethernet to fiber...

Страница 12: ...d COMMAND The backwards compatible Model 2090 Multi Device Controller command When used by itself controls the first device in a multi axis system and when arguments are required supports only a singl...

Страница 13: ...TS Lindgren Inc 2301 Precision Positioner Module Name PCA120518 FW N NN identifies this device as a 2301 Precision Positioner The Module Name parameter is a place holder to identify a specific module...

Страница 14: ...e Name parameter in the IDN query response is a place holder to identify a specific device in a network If you have more than one device you might want to identify them with different module names For...

Страница 15: ...e Command HOME Description The device has a mechanical home sensor Every time the positioner is turned on a home procedure must be performed so the current position is known by the firmware To home th...

Страница 16: ...pleted See Homing Procedure p 14 Example AXIS1 OPC Scan Command SCAN Description Instructs the positioner to begin scanning between preset lower and upper limits Example AXIS1 SCAN Seek Negative Comma...

Страница 17: ...nd lower counterclockwise limits Example AXIS2 SKP 90 Seek Relative Command SKR nnn n Description Instructs the device to begin seeking the specified target value relative to the current position The...

Страница 18: ...ample AXIS1 2 ST Trigger Configuration Command TRIGGER ON OFF step size reference pre trigger delay pulse length post trigger delay polarity Description Use this command to configure the trigger Where...

Страница 19: ...rection 5 Hardware Limit hit 6 Polarization limit violation 7 Lost communication 9 Encoder failure 10 Trigger failure 11 Motor overheat 12 Relay failure 13 Position out of bounds 32 Motor driver fault...

Страница 20: ...using the same IP address The IP address of the device will need to be changed if more than one device is in the same LAN To change the IP address of an ETS Lindgren Ethernet device use its embedded...

Страница 21: ...e the connection is working by typing IDN then pressing the Enter key on the keyboard The device will respond with an identification string such as the one shown below The IP address can be changed us...

Страница 22: ...22 Computer Network Configuration Connect to an ETS Lindgren Ethernet by setting the computer Ethernet interface to the selections shown below...

Страница 23: ...net cables It also communicates with the computer and allows it to access a network device Follow these steps on a Windows 10 PC to configure the Ethernet adapter 1 Ensure that the Ethernet adapter is...

Страница 24: ...ange Ethernet settings In the search results click change Ethernet settings In the Settings window click Change adapter options Right click on the Ethernet Adapter you intend to configure and select P...

Страница 25: ...192 168 0 1 Subnet mask 255 255 255 0 Default gateway blank Select Use the following DNS server addresses and leave Preferred and Alternate DNS server fields blank Then click OK Information about subn...

Страница 26: ......

Отзывы: