A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y
MyLab
- G E T T I N G S T A R T E D
2 - 4
Electromagnetic Immunity for All Medical Equipment
The
MyLab
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
MyLab
system should assure that it is used in such an environment.
Immunity Test
IEC60601
Test Level
Compliance
Level
Electromagnetic Environment
and Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
kV on contact
kV in air
kV on contact
kV in air
The floor should be in wood,
concrete or ceramic tiles. If floors
are covered with synthetic
material, the relative humidity
should be at least at 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines
2 kV for power
supply lines
1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
mode
1 kV differential
mode
2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips
IEC 61000-4-11
0%
U
T
; 0,5 cycles
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at
0°
0%
U
T
; 0,5 cycles
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at
0°
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the
MyLab
system requires
continued operation during
power mains interruptions, it is
recommended that the
MyLab
system is powered from an
uninterruptible power supply or a
battery.
Power interruptions
IEC 61000-4-11
0%
U
T
; 250/300
cycles
0%
U
T
; 250/300
cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the
MyLab
system requires
continued operation during
power mains interruptions, it is
recommended that the
MyLab
system is powered from an
uninterruptible power supply or a
battery.
Содержание MyLabX6
Страница 1: ...Rev 08 February 2020 MyLabX6 MyLabX7 GETTING STARTED 350031500...
Страница 2: ...MyLab G E T T I N G S T A R T E D ii...
Страница 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity...
Страница 6: ...MyLab G E T T I N G S T A R T E D vi...
Страница 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED RED Declaration of Conformity...
Страница 8: ...MyLab G E T T I N G S T A R T E D viii...
Страница 9: ...MyLab G E T T I N G S T A R T E D ix GETTING STARTED...
Страница 10: ...MyLab G E T T I N G S T A R T E D x...
Страница 16: ...I N D E X MyLab G E T T I N G S T A R T E D xvi...
Страница 28: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Страница 36: ...A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y MyLab G E T T I N G S T A R T E D 2 8...
Страница 49: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 13 GETTING STARTED...
Страница 66: ...P R E P A R I N G T H E S Y S T E M MyLab G E T T I N G S T A R T E D 4 12...