B2 WEIGHING INDICATOR
4.3
Weighing parameters
Press
key while “
PAr
“ message is on display. Weighing parameters are edited one at a time.
Edit each value by using
and
keys; press
to accept and proceed with the next
parameter. You may press
and go back to the previous parameter while editing. When all the
paremeters are accepted, the device saves the changes automatically.The parameters are:
1. Capacity 1
In single range instruments, this is the maximum capacity.
In multi (two) range instruments, this is the maximum capacity of the first
range (smaller range).
2. Capacity 2
In single range instruments, this value have to be entered as 0 (zero).
In multi (two) range instruments, this is the maximum capacity of the
second range (greater range).
3. Scale interval (e)
This value is the difference between two consecutive indicated values (e).
In multi range instruments, this is the interval of the smaller range (e1). The
device selects e2 automatically.
( Accepted values: 1,2,5,10,20,50 )
4. Decimal places
This value determines the place of the decimal dot while indicating weight
values.
( Accepted values: 0,1,2,3 )
5. Stability range
Stability of equilibrium is said to be reached, when the rate of change in displayed weight values
falls below a predetermined rate. This predetermined rate depends on 'Stability range' (SR) and
'Stability delay' (SD). If the difference of the maximum and the minimum weight values
calculated in a preiod of SD is less than or equal to SR, stability condition is satisfied.
The range for displayed weight values, for determination of the stability.
( Accepted values: 0 .. 5 interval(s) )
Note: A value of zero means 0.5 interval
v1.00 Page 11