B2 WEIGHING INDICATOR

4.4

Adjustment ( Calibration )

Press 

 key while “  

AyAr

 “ message is on display. There are 3 options for proceeding.

Dead load adjustment:

This option may be used when only dead load (zero point) adjustment is 
needed.

While this message is displayed, unload the platform and press 

 key 

when there is no motion. The display will become blank until the device 
calculates and saves the calibration data. After that, the device exits the 
menu and returns to the weight display.

Span adjustment:

This option may be used when a dead load adjustment is not needed, but a
span adjustment is needed. (The device assumes the last zero point for a 
reference for the adjustment weight.)

While this message is displayed, press 

 key. The dead load 

adjustment (zero calibration ) is skipped.

Enter (edit) the value of the weight which will be used for adjustment.
Accept with 

 key.

The value you entered will be displayed constantly. Load the platform with 
the adjustment weight you entered and press 

 key when there is no 

motion. The display will become blank until the device calculates and 
saves the calibration data. After that, the device exits the menu and returns
to the weight display.

Full calibration:

This option is performing the dead load adjustment and the span 
calibration consecutively.

While this message is displayed, unload the platform and press 

 key 

when there is no motion. The display will become blank until the device 
calculates and saves the calibration data.

Then Enter (edit) the value of the weight which will be used for calibration.
Accept with 

 key.

The value you entered will be displayed constantly. Load the platform with 
the calibration weight you entered and press 

 key when there is no 

motion. The display will become blank until the device calculates and 
saves the adjustment data. After that, the device exits the menu and 
returns to the weight display.

v1.00                                                                                                                                                       Page  13

USER MANUAL

SıFır

SıFır

015000

 15000

SıFır

015000

 15000

Содержание B2

Страница 1: ...B2 Weighing Indicator User Manual ...

Страница 2: ...1 Adjustment Calibration switch 8 4 2 Configuration menu 10 4 3 Weighing parameters 11 4 4 Adjustment Calibration 13 4 5 Miscellaneous Application Settings 14 4 6 Serial Output Settings 15 5 Operation 17 5 1 Weighing with tare 17 5 2 High resolution display 17 5 3 Multi range weighing 17 6 Appendix 18 6 1 Error messages 18 6 2 Specifications 19 v1 00 Page 2 USER MANUAL ...

Страница 3: ...ccurate weighing Internal alibi memory for recording weighing results RS 232 ports for connections to printers remote displays computers or other digital terminal equipments LED display This manual contains the information needed to configure adjust and operate the device To ensure safe operation the instructions in this manual shall be respected v1 00 Page 3 USER MANUAL ...

Страница 4: ...n is 1 mm If the cable length is too long use multiple wires for Excitation connections If the sense inputs on the load cell connector are left empty the device will display Err 3 Make sure all the pins are connected Shielded twisted pair cables must be used for load cell and communication For best accuracy the device shall be powered on at least 30 minutes before adjustment Device setup shall be ...

Страница 5: ...dicator Indicates that a tare device is in operation and the displayed weight value is a net weight Stability No motion indicator Indicates that stability of equilibrium has been reached kg Unit of weight Indicates the unit of the weight if a weight value is displayed BAT Low battery indicator If flashing indicates a low battery condition F Function indicator Indicates that a function is in operat...

Страница 6: ...elled and gross weight is indicated Print key If the indicated value is positive and stable it is saved as a weighing result and a print out is produced if configured Star key Call up the configuration menu Please refer to the configuration section Function key Activates and deactivates a configured function For a complete explanation refer to the related section of this manual Note For an explana...

Страница 7: ... is 4 wire connect Excitation here 4 Sense Note If load cell is 4 wire connect Excitation here 5 Ground Shield 6 Excitation 7 Excitation USB RS 232 Pin Signal PC Connection Type D 9 Pin Male 2 Tx 2 Rx 3 Rx 3 Tx 7 Gnd 5 Gnd Remote Display Pin Signal ERTE HG57 Remote Display Connection Type D 9 Pin Female 1 2 3 NC No connection 4 5 6 7 8 9 4 5 6 7 8 9 one to one v1 00 Page 7 USER MANUAL ...

Страница 8: ...ermines the mode of configuration If the switch is in LOCKED position it is not possible to alter weighing related settings If the switch is in UNLOCKED position it is possible to edit weighing parameters and adjustment data v1 00 Page 8 USER MANUAL Main board Display board Keypad board Connectors INTERNAL VIEW ...

Страница 9: ...IGHING INDICATOR v1 00 Page 9 USER MANUAL ON OFF ADJUSTMENT SWITCH 1 2 ON OFF 1 2 In this position adjustment is OFF LOCKED In this position adjustment is ON UNLOCKED Note Position of switch 2 has no effect ...

Страница 10: ...el Return to a higher menu level When user input is needed while changing a device setting you are prompted to edit a numerical value For this purpose a numerical value is displayed with a flashing digit During the edit procedure the function of the keys are Clear The whole value is reset to zero Next digit The flashing moves to the next digit at right If the rightmost digit is flashing the flashi...

Страница 11: ...e 3 Scale interval e This value is the difference between two consecutive indicated values e In multi range instruments this is the interval of the smaller range e1 The device selects e2 automatically Accepted values 1 2 5 10 20 50 4 Decimal places This value determines the place of the decimal dot while indicating weight values Accepted values 0 1 2 3 5 Stability range Stability of equilibrium is...

Страница 12: ...anges in weight A lower value causes fast reactions to weight changes and is suitable in relatively stable conditions A higher value causes slow reactions but the vibrational flickers are eliminated Accepted values 1 5 8 Zero at power on This setting determines if the indication is set to zero automatically at power on initial zero setting Accepted values 0 off 1 on v1 00 Page 12 USER MANUAL hc 1 ...

Страница 13: ...value you entered will be displayed constantly Load the platform with the adjustment weight you entered and press key when there is no motion The display will become blank until the device calculates and saves the calibration data After that the device exits the menu and returns to the weight display Full calibration This option is performing the dead load adjustment and the span calibration conse...

Страница 14: ...Fon message is on display Automatic Tare 0 Off 1 On Automatic Tare Cancel 0 Off 1 On Automatic Printing This option is only for indicators with integrated mini printer 0 Off 1 On Automatic Tare Limit e 0 Off or of e divisions Automatic Printing Limit e This option is only for indicators with integrated mini printer 0 Off or of e divisions v1 00 Page 14 USER MANUAL od 0 di 0 oy 0 dL 0 yL 0 ...

Страница 15: ... to the previous parameter while editing When all the paremeters are accepted the device saves the changes automatically RS232 Output Mode 0 Continuous Output Mode 1 Terminal Mode See below for details on Continuous and Terminal Mode protocols Baud Rate 0 9600 1 4800 2 2400 3 1200 4 19200 Output Mode Type If you don t need different serial output format don t change the defaults 0 ERTE B2 series o...

Страница 16: ...mal point no unit e g 1500 8 character Carriage Return ASCII 13 Terminal Mode Output Format 9600 baud 8 data bit 1 stop bit no parity 9600 8 N 1 In this mode the indicator responds to specific commands via the serial interface The commands and the responses are as below Command Response d Sends the indication value as seen in the display in the form of continuous mode protocol z Zero Key No respon...

Страница 17: ...d press the key Unstable and negative weight values can not be tared 5 2 High resolution display This function needs the function key configured for high resolution display It is possible to temporarily display measured weight values in 10 fold increased resolution To perform this function press the key The indicated weight value is displayed in high resolution for 5 seconds At the end of this per...

Страница 18: ...error Possible disconnection or erroneous connection of load cell cable Check load cell connection hata 4 Initial zero setting error Weight value is above or below zero setting range Check the platform hata 92 when performing an adjustment Calculations show that load cell signal may be out of range for weighings near capacity max The adjustment will be accepted this is just a warning hata 99 when ...

Страница 19: ...lications 40000e Minimum input voltage per verification scale interval 0 8μV e approved 0 2μV e industrial Input Signal Range 0 20 mV Measurement Method Delta sigma A D 20ms 1s update rate Internal Resolution 6 700 000 adım Linearity 0 007 Interfaces RS 232 serial port USB connection PC ERTE HG External Display Environmental Operating Temperature 10 C 40 C Storage Temperature 30 C 80 C Electromagn...

Отзывы: