Captair Bio Smart
Mobile ductless filtering PCR workstations
16
Replacing the filters
Your device is equipped with FLEX™ filter technology that was configured to the user’s protection needs when the device
was purchased. The configuration of the filter column is dependent on the applications carried out in the enclosure. These
applications may change over time. Your FLEX™ filter technology can therefore be reconfigured if your workstation is used
for applications other than those anticipated when the device is first set up. If so, please contact us so that we can verify that
the current configuration is safe or it needs to be reconfigured.
The table below summarises all possible Flex™ filter technology configurations for your device
Captair 321 - 391 - 481 Smart models
Molecular filter
HEPA filter H14 / ULPA
Pre-filter
Column
Configura-
tion
1C
x1
x1
1C1P
x1
x1
x1
UP
The table below summarises the different types of carbon filters that Erlab® offers along with their fields of application.
Type AS
For organic vapors
HEPA H14
For particles
ULPA U16
For particles