950-109 rev A
ergonomic products • 198 Airport Rd • Fall River, MA 02720 • 866-374-6487
PAGE 21
Safe-T-Shield™ VacTrac
Model T2B1
WEEKLY CLEANING AND DISINFECTION
The Suction Tube Interior
In addition to cleaning and disinfection of the exterior elements of the Safe-T-Shield after each use, we
recommend disinfecting the interior of the Suction Tube weekly.
The best method for this is to employ a fogger using hypo-chlorous acid.
1. Remove the Shield from the manifold (see p. 22).
2. Remove the stainless steel Airflow Plate from the manifold (see p. 23).
3. Follow your manufacturer’s instructions to prepare your fogger for use.
4. Turn on the Safe-T-Shield.
5. Place the fogger nozzle in the center of the Suction Tube
6. Fog the tube for 10 seconds only and stop. To avoid saturation of the Primary Cotton Filter, do not
run the fogger longer than the recommended 10 seconds.
7. Turn off the Safe-T-Shield and replace the Airflow Plate and Shield (see pp. 22 & 23).
The Airflow Plate
The Airflow Plate may be autoclaved. We recommend this weekly as well. See page 23 for instructions
on removal and re-installation of the Plate.
As previously mentioned, some clients prefer to have spare Airflow Plates available for a quick swap
out should they require additional cleaning to maintain a pristine appearance.
9 - CLEANING AND DISINFECTION cont.