4  Preparing the Patient

4  Preparing the Patient

4.1  Handlebar Adjustment

To adjust the handlebar angle, open the rotary lever 

1

 by 

turning it counter‑clockwise.

Choose a handlebar angle that allows the patient to sit up 
straight and comfortably. Then tighten the rotary lever 

1

 

hand tight by turning it clockwise.

Before allowing the patient to rest the full body weight on 
the handlebar, check the clamping as follows:

1

Figure 4 – 1:   Handlebar  adjustment 

1

  Rotary lever

Danger

•  With the ergometer standing firmly, check that 

the handlebar is tight by trying to push the 
handlebar downwards from above. Adjust the 
clamping force of the rotary lever if necessary.

The handlebar is not designed to support the full body weight! 
Risk of falling!

4.2  Saddle adjustment

The saddle height of the ergoselect 1 is adjusted manually with a 
rotary lever.

When the pedal is in its lower position, there should be a 
10° angle between the axis formed by the upper body and 
the thigh.

Set the handlebar to a position where it is comfortable to 
reach for the patient while sitting upright (see section 4.1 

Handlebar Adjustment

 on page 10).

Figure 4 – 2:  Adjusting Saddle and Handlebar

1

  Adjusting the handlebar angle

2

  Adjusting the height of the saddle

Содержание ergoselect 1

Страница 1: ...ergoselect 1 Bicycle Ergometer Operator s Manual 201000551000 Version 2021 05 12 Rev 02 English...

Страница 2: ......

Страница 3: ...n is acknowledged No part of this manual may be reprinted translated or reproduced without the manufacturer s written permission This manual will not be automatically updated Please contact the manufa...

Страница 4: ...2 ergoselect 1...

Страница 5: ...Indicators 12 5 2 Mounting the Control Terminal 12 5 3 Transport 13 5 4 Setup 13 5 5 Connecting the Power Cord 14 5 6 Connecting the ECG Cable 15 6 Operation 16 6 1 Turning the System On 16 6 2 Operat...

Страница 6: ...e 27 9 2 Technical Safety Inspections Inspections of the Measuring System 27 9 3 Disposal 27 10 Technical Specifications 28 10 1 Ergometer 28 10 2 Exercise Test Protocols 29 10 3 Family of characteris...

Страница 7: ...he device All device users and persons responsible for assembly maintenance inspection and repair of the device must read and understand the content of this manual before using the device or working w...

Страница 8: ...s may not be used in the immediate vicinity of the ergometer because they might interfere with the proper functioning of the ergometer Electromagnetic interference most probably exists when the watt r...

Страница 9: ...the device dizziness nausea or pain patients under the influence of substances that may impair vigilance alcohol drugs medication Contraindications in exercise testing carried out with ergometers sour...

Страница 10: ...ed specialists The group of intended users does not include persons whose mental and physical capabilities and skills have an adverse effect on their ability to use the medical device in accordance wi...

Страница 11: ...ion e g March 2020 Toggle switch ON voltage Toggle switch OFF voltage CE mark per the Medical Device Directive 93 42 EEC of the European Union Notified body T V S D Product Service GmbH Ridlerstr 65 8...

Страница 12: ...ing firmly check that the handlebar is tight by trying to push the handlebar downwards from above Adjust the clamping force of the rotary lever if necessary The handlebar is not designed to support th...

Страница 13: ...by turning it clockwise Figure 4 3 Tightening the rotary lever Warning Do not choose saddle height settings above the max mark Do not exceed the maximum height marked on the scale to avoid any risk of...

Страница 14: ...st the ergometer to uneven floors 8 Sockets for power cord and connection cables under side of ergometer Figure 5 1 Operating controls of the ergoselect 1 5 2 Mounting the Control Terminal The control...

Страница 15: ...t the ergoselect 1 When you have reached the new location lower the ergoselect 1 very carefully to protect it from considerable damage Caution Equipment Damage Avoid strong vibrations of the ergoselec...

Страница 16: ...the handlebar Plug the power cord into a wall outlet Figure 5 6 Power cord in strain relief mounted to frame Warning Arrange the power cord properly Lay the power cord flat on the floor Keep the powe...

Страница 17: ...ed into the 9 pole socket of the connection panel Port 1 or the USB port and secured at the metal frame with an additional strain relief Figure 5 7 Connection for ECG recorder PC ECG system 1 USB PC c...

Страница 18: ...Selftest running Figure 6 2 Self test screen Note Instructthepatientnottopedalwhiletheergometer is being turned on and during the self test The device can be configured to default to one of the operat...

Страница 19: ...n automatic exercise test some of the corresponding test protocols are user configurable and stored in the system see chapter 10 2 Exercise Test Protocols on page 29 Manual The ergometer is controlled...

Страница 20: ...ter receives commands from the controlling ECG unit or PC the exercise test will start and the corresponding values will be displayed The exercise test can only be terminated with the corre sponding c...

Страница 21: ...the display after approx 15 seconds and the patient s RPM indicator blinks the patient should start pedalling 0 0 Watt min sec min Start Figure 6 11 Starting the exercise test The internal protocol w...

Страница 22: ...tarted with the Start key afterwards the load can be set and changed with the 10 W and 10 W keys in increments of 1 W 5 W 10 W or 25 W as configured 0 0 Watt min sec min Start Figure 6 15 Initial scre...

Страница 23: ...s menu 6 9 1 Default Mode In this menu you choose the default mode activated when the ergometer is turned on When first turned on after delivery the ergometer will display this menu Use the softkeys o...

Страница 24: ...e the new value The other parameters are edited in the same way You exit the configuration with Protocol 6 Select Step Basic Load 20 W Stage Time min Load Stage W Select Figure 6 22 Editing the parame...

Страница 25: ...ental change of this setting the menu is protected with a password Enter the service code 003 and confirm with Select Select ECG Type Service Code 003 Figure 6 27 Entering the EKG Type password The er...

Страница 26: ...in Max 50 130 65 min Select Figure 6 29 Setting the RPM limit values 6 9 9 HR Belt Number If the test subject wears a chest strap to measure the heart rate during the training the corresponding chest...

Страница 27: ...s adjustable Ergonomics optional erg705 942 Pedal cranks adjustable w o tools Ergonomics optional erg705 786 Pedal extra wide with Comfort pedal straps set Pedal optional erg705 300 Tilt protection St...

Страница 28: ...isinfection and General Hygiene Measures The document Cleaning and Disinfecting ergoline Medical Devices Part No 201000641000 in its most recent version is also part of this manual This document is ex...

Страница 29: ...the measuring system must be completed every two years according to the rules of the art by a Service Engineer authorized by ergoline GmbH The date of the next inspection is indicated on the inspec t...

Страница 30: ...ols Control Terminal P 5 predefined incremental protocols WHO Hollmann etc 10 user programmable exercise test protocols Permitted patient weight 160 kg max Permitted patient height approx 120 210 cm c...

Страница 31: ...nment operation temperature 10 to 40 C rel humidity 30 to 75 no condensation atmospheric pressure 800 to 1060 hPa transport and storage temperature 20 to 70 C rel humidity 10 to 95 no condensation atm...

Страница 32: ...20 20 20 20 20 20 20 20 20 20 20 20 20 20 35 min 10 35 min 10 35 min 10 20 50 min 20 50 min 20 50 min 70 min 40 min 20 70 min 20 70 min 30min 10 30 min 10 30 min 10 30 min 10 30 min 10 30 min 10 30 mi...

Страница 33: ...system should be tested to verify normal operation in the configuration in which it is being used Guidance and Manufacturer s Declaration Electromagnetic Emissions Theergoselect1ergometerisintendedfor...

Страница 34: ...be that of a typical commercial or hospital environment Surge to EN 61000 4 5 1 kV differential mode 2 kV common mode 1 kV N A Mains power should be that of a typical commercial or hospital environme...

Страница 35: ...distance in meters m Field strengths from fixed RF transmit ters as determined by an electromag netic site surveya should be less than the compliance level in each frequency rangeb Interference may o...

Страница 36: ...power of the communications equipment Rated Maximum Output Power of Transmitter W Separation Distance according to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 M...

Страница 37: ......

Страница 38: ...ergoline GmbH Lindenstra e 5 72475 Bitz Germany Tel 49 0 7431 98 94 0 Fax 49 0 7431 98 94 128 e mail info ergoline com http www ergoline com...

Отзывы: