Electromagnetic interference immunity, compliance and test level
Electromagnetic Inter-
ference immunity
Measurements
Required
Observed
Static electricity discharge
(ESD) in compliance with
IEC 61000-4-2
Contact ±8 kV
Air ±2 kV, ±4 kV, ±8 kV,
±15 kV
Contact ±8 kV
Air ±15 kV
RF radiation in compliance
with IEC 61000-4-3
3 V/m or 10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
RF radiation in the imme-
diate vicinity of wireless
communication devices
in compliance with IEC
61000-4-3
see the following table
see the following table
Magnetic filed for supply
frequency (50/60 Hz)
in compliance with IEC
61000-4-8
30 A/m
50Hz or 60Hz
100 A/m
50 Hz
Fast transient electrical
disturbances/bursts in com-
pliance with IEC 61000-4-4
+/- 2 kV / 100 kHz repeti-
tion frequency for power
cable
+/- 2 kV / 100 kHz
repetition frequency for
power cable
Surges in compliance with
IEC 61000-4-5
Line - Line: ±0.5 kV, ±1 kV
Line - PE: ±0.5 kV, ±1 kV,
±2 kV
Line - Line: ±0.5 kV, ±1
kV Line - PE: ±0.5 kV, ±1
kV, ±2 kV
Conducted RF interference
in compliance with IEC
6100-4-6
6 Vrms
150 kHz to 80 MHz
6 Vrms
150 kHz to 80 MHz
Voltage dips, brief voltage
interruptions and voltage
fluctuation in compliance
with IEC 61000-4-11
30 % 10ms
→
B
60 % 100 ms
→
C
>98 %5000ms
→
C
30 % 10ms
→
A
60 % 100 ms
→
A
>98 %5,000ms
→
A
9 1