23
4.4 Disconnecting the product from the unit
1. Lateral forces can damage the receptacle. Therefore ensure
you pull the HF connector of the adapter out of the multi-
functional receptacle in a straight line.
2. Unscrew the argon gas connection of the adapter and pull it
out of the argon gas receptacle in a straight line.
5
Cleaning, disinfection, sterilization
5.1 Safety
instructions
Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Overview of validated
procedures“.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
OBJ_DOKU-194452-003.fm Seite 23 Donnerstag, 2. April 2020 2:47 14
Содержание 20132-249
Страница 2: ...OBJ_DOKU 194452 003 fm Seite 2 Donnerstag 2 April 2020 2 47 14 ...
Страница 30: ...30 OBJ_DOKU 194452 003 fm Seite 30 Donnerstag 2 April 2020 2 47 14 ...
Страница 102: ...102 OBJ_DOKU 194452 003 fm Seite 102 Donnerstag 2 April 2020 2 47 14 ...
Страница 216: ...216 OBJ_DOKU 194452 003 fm Seite 216 Donnerstag 2 April 2020 2 47 14 ...
Страница 243: ...KO 사용 설명서 APC 소켓 장착 APC 2 용 APC 어플 리케이터 어댑터 20132 249 OBJ_DOKU 194452 003 fm Seite 243 Donnerstag 2 April 2020 2 47 14 ...
Страница 256: ...256 OBJ_DOKU 194452 003 fm Seite 256 Donnerstag 2 April 2020 2 47 14 ...
Страница 257: ...OBJ_DOKU 194452 003 fm Seite 257 Donnerstag 2 April 2020 2 47 14 ...