22
4.3 Connecting the product
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the adapter.
IMPORTANT!
The adapter is only suitable for connecting argon
plasma applicators. APC probes must not be connected. If APC
probes are connected to the adapter, however, a message ap-
pears on the device which must be observed.
1. Switch on the VIO electrosurgical unit and select the basic
program.
2. Insert the HF connector of the adapter (1) into the multifunc-
tional receptacle of the APC 2. The connector must click into
place audibly. A message appears on the unit that can be
skipped and that disappears as soon as an argon plasma ap-
plicator is connected to the adapter.
3. Screw the argon gas connection of the adapter (2) into the
argon gas receptacle of the APC 2.
4. When the adapter is connected, a standard setting for mode,
power, effect and gas flow is automatically made on the VIO
electrosurgical unit. However, the settings can be modified
within a certain range if required.
2
1
OBJ_DOKU-194452-003.fm Seite 22 Donnerstag, 2. April 2020 2:47 14
Содержание 20132-249
Страница 2: ...OBJ_DOKU 194452 003 fm Seite 2 Donnerstag 2 April 2020 2 47 14 ...
Страница 30: ...30 OBJ_DOKU 194452 003 fm Seite 30 Donnerstag 2 April 2020 2 47 14 ...
Страница 102: ...102 OBJ_DOKU 194452 003 fm Seite 102 Donnerstag 2 April 2020 2 47 14 ...
Страница 216: ...216 OBJ_DOKU 194452 003 fm Seite 216 Donnerstag 2 April 2020 2 47 14 ...
Страница 243: ...KO 사용 설명서 APC 소켓 장착 APC 2 용 APC 어플 리케이터 어댑터 20132 249 OBJ_DOKU 194452 003 fm Seite 243 Donnerstag 2 April 2020 2 47 14 ...
Страница 256: ...256 OBJ_DOKU 194452 003 fm Seite 256 Donnerstag 2 April 2020 2 47 14 ...
Страница 257: ...OBJ_DOKU 194452 003 fm Seite 257 Donnerstag 2 April 2020 2 47 14 ...