c
Important:
❏
Never touch the white calibration tile on the inside of the sensor cover. Doing so could affect the color
measurement accuracy.
❏
Do not use volatile chemicals such as alcohol, thinner, or neutral detergents or liquid such as water. These
substances may affect the performance of the device.
❏
Never disassemble the device in any way other than the following procedure.
Note:
After performing color measurement, make sure you close the sensor cover on the base of the device to prevent dust and so on
from entering the sensor.
A
Confirm that the device is off.
Disconnect the USB cable if it is connected.
B
Open the sensor cover.
C
Loosen the screw, and then remove the plastic part.
Use a precision screwdriver that matches the size of the cross-hole in the screw head.
SD-10 User's Guide
Maintenance
37