Environmental Instruments Fluoroprobe FL-2-S0 Скачать руководство пользователя страница 6

 
 

2. Zero Calibration:

 First prepare a container of oxygen depleted water.  One of the 

chemicals that are effective in depleting the 0

2

 in water is Sodium Sulfite (NaSO

2

).  Use 

one gram/quart of water, mix well and allow up to one hour to deplete the oxygen before 
using.  

Warning: let probe sit in prepared water for (30) minutes or longer in order 

to allow the sensor to equilibrate prior to adjustment.  

It may take as long as (45) 

minutes or more to go from air to zero level, so make sure a stable zero level has been 
obtained before adjusting the instrument.  The probe can be gently moved up or down 
to speed the process, however no flow is required to calibrate or operate the instrument. 
When ready, press the red read button on the faceplate of the unit.  This will allow (2) 
minutes to adjust the instrument using the zero calibrate set point.  After adjustment, the 
unit will be calibrated to zero (0).  Changing the probe from zero to air can contaminate 
the calibration solution so be sure to rinse the probe clean before each step. 

 

3. Air Calibration:

 Air calibration is an easy and reliable method to use to calibrate the 

unit.  This technique must be performed out of direct light.  If the probe is mounted in a 
float device for aquaculture or mounted in a pipe for wastewater, pull the probe out of 
the water and wipe off the sensor end.  After making sure the sensor is clean. 

Note:

 If 

the end of the probe is dried cleaned with a soft toothbrush the sensor will air saturate 
quicker. To calibrate, place the probe back in the pipe and leave probe suspended 
between the holes above the water for about twenty (20) minutes. Do not air calibrate if 
air and water temperatures are + or – 10

o

C difference. 

 

Using the enclosed pressure/altitude chart (fig-3) and the altitude of the location where 
the unit will be housed, select the proper calibration value.  Press the red 

Read

 button 

on the faceplate of the unit.  The operator will have two (2) minutes to adjust the Air 
calibrate pot using a small screwdriver.  After completing the adjustment, replace the 
probe back in to the water.  

Warning: If the pipe is submerged it has to be vented 

below the point at which the probe is in the air due to treatment gases in the pipe. 

 

4. Calibration using hand held DO meter:

 Take handheld readings in the location where 

the Fluoroprobe

TM

 has been installed.  As soon as the hand held reading has been 

taken, make sure the selector switch is on DO, then press the red 

READ

 button.  The 

operator will have two ( 2 ) minutes to make the adjustment to the unit.  Allow the 
reading to stabilize before making adjustments.  Uses the Air calibrate pot for 4ppm and 
greater or use the zero for 4ppm and below.  Adjust to the handheld DO meter reading. 
The instrument is now ready to operate and needs no further adjustment.

 

 
Warning:. If the unit has just been put in the water, allow at least two (2 or 3) days 
to equilibrate to the type of environment the sensor will be used in before 
calibrating.  Do not calibrate if probe has been removed from water, wait 30 
minutes or longer before adjusting.  

 

5. Probe and Sensor Maintenance:

 The factory request that the probe be inspected at 

least once a month.  The probe can be washed down in the probe holder before the 
inspection of the sensor takes place.  A soft toothbrush should be used on the black 
sensor on the end of the probe.  Place back in water and allow at least (1) one hour to 
stabilize. 

Содержание Fluoroprobe FL-2-S0

Страница 1: ...O M Manual MODEL FL 2 S0 NEW WEBSITE WWW thefluoroprobe com ENVIRONMENTAL INSTRUMENTS LLC 144 Industrial Park Drive SuiteA 3 Waynesville NC 28786 From Process Control Services Phone 02 9319 1808 Fax...

Страница 2: ...ibration 3 Air Calibration 4 Calibration Using Hand Held DO Meter 5 Probe and Sensor Maintenance 6 Troubleshooting and Tips Figure 1 Installation Illustration Figure 2 Face Plate Descriptions Drawing...

Страница 3: ...tional solar powered brush for cleaning the sensor can be ordered mounted on a float for remote or aquaculture applications Unlike polargraphic and Galvanic probes the FluoroprobeTM sensor does not co...

Страница 4: ...ibrating The FluoroprobeTM is very stable and the longer it stays in the water the better The FluoroprobeTM is calibrated before shipping to a known zero 0 ppm standard and air saturated water 8 47 D...

Страница 5: ...owth or debris covering the sensor will cause oxygen reading to slowly fall to zero the only maintenance needed is once a month toothbrushing Many aeration basins never have this growth problem 1 Air...

Страница 6: ...t twenty 20 minutes Do not air calibrate if air and water temperatures are or 10o C difference Using the enclosed pressure altitude chart fig 3 and the altitude of the location where the unit will be...

Страница 7: ......

Страница 8: ......

Страница 9: ...1 2 3 4 1 2 3 4 A B C D A B C D Title Size Number Rev Date File Name Drawn By Sheet of RELAY LOCATION RELAY 3 10A 250VAC RELAY 2 10A 250VAC RELAY 15A 250VAC TERMINAL POS WARRANTY ZERO LEAVE ALONE...

Страница 10: ...ile Name Drawn By Sheet of THIS POT ADJUSTS THE 4mA AT 0 PPM THIS POT ADJUSTS THE SPAN CAN BE ADJUSTED FOR 20mA AT 10 PPM TO 19 99 PPM BLUE WIRE 4 20 WHITE WIRE 4 20 NOTE MAX LOOP RESISTANCE IS 550 oh...

Страница 11: ...MODEL FL 2 SWING OUT WIRING DIAGRAM LINE NEU 4 20 4 20 Earth RELAY 1 RELAY 1 RELAY 2 RELAY 2 230 VAC RELAY 3 RELAY 3...

Страница 12: ...0 2ppm Sensor Lifetime Based solely upon the application environment and the effect on Silicone Rubber 3 year Sensor warranty on approved applications 4 20 mA Max Resistance 550 ohms Features Tempera...

Страница 13: ......

Отзывы: