Environmental Instruments Fluoroprobe FL-2-S0 Скачать руководство пользователя страница 4

Controls and Set Points 

 

HIGH SET: 

Turn the rotary switch to 

HIGH SET 

and adjust the 

HIGH SET 

pot (on the right 

side of the controller) to the level you select as a cut off point.  This adjustment can be 
made with a small screwdriver.  This set point must be Higher than the 

LOW SET 

in order 

for the Low- High relay to operate. 
 

LOW SET

: Turn the rotary switch to 

LOW SET

 and adjust the 

LOW SET

 pot (on the lower 

right corner of the controller) to the level you select as the first relay cut on point.  This 
adjustment can he made with a small screwdriver.  This set point must be Lower than the 

HIGH SET

 in order for the Low-High relay to operate.  

 

SET 2

 is a back up relay for auxiliary equipment in case the oxygen level keeps falling.  

This relay has one setting, the 

HIGH SET

 will act as the cutoff. 

 

SET 3

 operates the same as 

SET 2

 and can be used for back up if the oxygen levels keep 

decreasing or an alarm. 
 

OPERATION

: Turn the rotary switch to the 

DO

 position and the system will operate 

automatically in accordance with the setting you have selected.  

NOTE: If 4-20 mA output 

is being monitored by a computer or data logger, etc. The rotary switch must be in 
the D.O. position.  The relay set points are not affected by the position of the rotary 
switch. 

 

Section 3 Introduction to Calibration  

 

After installing the unit, allow the sensor 2 or 3 days in its new environment to stabilize 
before calibrating.  The Fluoroprobe

TM

 is very stable and the longer it stays in the water, 

the better.  The Fluoroprobe

TM

 is calibrated before shipping to a known zero (0) ppm 

standard, and air saturated water (8.47 D.0. @ 20

o

C / 3000ft). Generally, the Fluoroprobe 

will not require calibration after installation, however, it is recommended to check the 
calibration at first use. 
 
The operator should take D.O. readings with their hand held probe over the first few days 
at the location where the probe is installed.  It is important that several days of readings 
with your handheld be observed before you make any adjustments to the Fluoroprobe

TM

The best thing to do if the units' readings appear odd is to do nothing and investigate 
thoroughly before making any adjustments.  If your operation range is from 4.0 ppm down 
to 0.0 ppm, you should calibrate to a portable meter or a know standard in that range. 
Always use the 

ZERO

 calibration pot to make adjustments below 4.0 ppm.  When the 

range is from 4.0 ppm and higher or you are calibrating to ambient air, use the 

AIR

 

calibration pot only. 
 

Note-. If air calibrating, allow thirty (30) minutes or more to stabilize the unit before 
comparing readings or making any adjustments.

  If the FL-2 reads zero (0.0) ppm, 

check the oxygen level with a handheld and then check the probe for trash blockage 
and/or clean the probe sensor with a small toothbrush.  If the display stays at 19.90 all the 
time the unit has other problems and should be sent back to the manufacturer.  

You may 

give us a call anytime for help - please see the back of the manual for the telephone 
number. 

Содержание Fluoroprobe FL-2-S0

Страница 1: ...O M Manual MODEL FL 2 S0 NEW WEBSITE WWW thefluoroprobe com ENVIRONMENTAL INSTRUMENTS LLC 144 Industrial Park Drive SuiteA 3 Waynesville NC 28786 From Process Control Services Phone 02 9319 1808 Fax...

Страница 2: ...ibration 3 Air Calibration 4 Calibration Using Hand Held DO Meter 5 Probe and Sensor Maintenance 6 Troubleshooting and Tips Figure 1 Installation Illustration Figure 2 Face Plate Descriptions Drawing...

Страница 3: ...tional solar powered brush for cleaning the sensor can be ordered mounted on a float for remote or aquaculture applications Unlike polargraphic and Galvanic probes the FluoroprobeTM sensor does not co...

Страница 4: ...ibrating The FluoroprobeTM is very stable and the longer it stays in the water the better The FluoroprobeTM is calibrated before shipping to a known zero 0 ppm standard and air saturated water 8 47 D...

Страница 5: ...owth or debris covering the sensor will cause oxygen reading to slowly fall to zero the only maintenance needed is once a month toothbrushing Many aeration basins never have this growth problem 1 Air...

Страница 6: ...t twenty 20 minutes Do not air calibrate if air and water temperatures are or 10o C difference Using the enclosed pressure altitude chart fig 3 and the altitude of the location where the unit will be...

Страница 7: ......

Страница 8: ......

Страница 9: ...1 2 3 4 1 2 3 4 A B C D A B C D Title Size Number Rev Date File Name Drawn By Sheet of RELAY LOCATION RELAY 3 10A 250VAC RELAY 2 10A 250VAC RELAY 15A 250VAC TERMINAL POS WARRANTY ZERO LEAVE ALONE...

Страница 10: ...ile Name Drawn By Sheet of THIS POT ADJUSTS THE 4mA AT 0 PPM THIS POT ADJUSTS THE SPAN CAN BE ADJUSTED FOR 20mA AT 10 PPM TO 19 99 PPM BLUE WIRE 4 20 WHITE WIRE 4 20 NOTE MAX LOOP RESISTANCE IS 550 oh...

Страница 11: ...MODEL FL 2 SWING OUT WIRING DIAGRAM LINE NEU 4 20 4 20 Earth RELAY 1 RELAY 1 RELAY 2 RELAY 2 230 VAC RELAY 3 RELAY 3...

Страница 12: ...0 2ppm Sensor Lifetime Based solely upon the application environment and the effect on Silicone Rubber 3 year Sensor warranty on approved applications 4 20 mA Max Resistance 550 ohms Features Tempera...

Страница 13: ......

Отзывы: