Error message
45
E2100
Sensor ADC error
The ADC system of Sensor
has a problem.
Remove the batteries,
disconnect the AC power
supply, and then press the
power key.
E2101
Sensor memory
error
The memory system of
Sensor has a problem.
E22XX
Sensor
communication
error
The communication system
of Sensor has a problem.
Disconnect the sensor
connector from meter, and then
reconnect the sensor connector
to the meter. Or, contact your
dealer for repair.
E2250
Sensor update
request error
Sensor update did not
complete successfully.
Execute sensor update again.
E2251
Sensor update
error
Error occurred during
sensor update.
E2300
Calibration
interval alarm
error
Exceeds the calibration
interval setup.
Calibrate the meter.
E2301
CAL error:
Offset
The offset potential of the
pH sensor is out of the
range of ±50 mV.
Maintenance the sensor or use
the new standard solution.
E2302
CAL error:
Slope
The slope of sensor is out
of range.
E2303
CAL error:
CAL points
Maximum calibration points
exceeded.
Up to maximum calibration
points.
E2304
CAL error:
Unknown solution
The meter cannot
recognize the standard
solution.
(1) Make sure to place the
sensor in the standard
solution.
(2) Make sure the buffer group
setting.
(3) Clean the sensor.
(4) Replace the new sensor.
E2305
CAL error:
Cell constant
Cell constant is out of
range.
(1) Make sure to place the
sensor in the standard
solution.
(2) Make sure the buffer group
setting.
(3) Clean the sensor.
(4) Replace the new sensor.
E2306
CAL error:
Temp. CAL
Temperature set for
calibration is more than
±2°C.
Make sure to place the sensor in
the standard solution.
E2308
CAL error:
Membrane
DO membrane has reached
the lifetime.
Replace with a new sensor cap.
Error
code
Error
description
Cause of error
Recommended Action
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30
Содержание Oakton 300 Series
Страница 1: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 4: ...Regulations III www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 11: ...Contents X www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 24: ...Viewing Calibration data 13 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 62: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...