Temperature calibration
22
Temperature calibration
Temperature calibration is required to accurately match the sensor to the meter. Check
the temperature reading and if its acceptable, no temperature calibration is required. If
you need to calibrate, please follow the procedure detailed below:
Prerequisites
Clean the sensor with DI (deionized) water and wipe it with tissue paper.
Switch on the meter and plug in the sensor.
Prepare standard solution required for calibration.
Set the Focus area in the measurement mode that the temperature is needed.
Place the sensor at least 3 cm in the standard solution.
Wait for 5 minutes to ensure temperature stability.
Note
In the only ATC setting, temperature calibration is available.
Temperature calibration must be performed using a known temperature solution or
against a calibrated thermometer.
Tip
To abort an ongoing calibration process at any point of time, press the
key.
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30
Содержание Oakton 300 Series
Страница 1: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 4: ...Regulations III www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 11: ...Contents X www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 24: ...Viewing Calibration data 13 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 62: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...