pH Setup
35
pH Setup
When the pH sensor head is connected to the meter, it is possible to change settings
suitable for measurement and calibration. In the measurement screen mode, move the
focus in the pH measurement item, and press the
key .
Default settings are in
Bold
.
*1: DIN19267
Parameter
Settings
Details
pH Setup
Use pH Setup to customize settings for the pH
sensor head.
pH Resolution
0.01
0.001
Set the resolution of the pH measurement value.
Buffer Group
Set the buffer type for automatic buffer recognition
during pH calibrations.
USA
USA:1.68, 4.01, 7.00, 10.01 and 12.45 at 25°C
DIN
*1
DIN:1.09, 3.06, 4.65, 6.79, 9.23 and 12.75 at 25°C
NIST
NIST:1.68, 4.01, 6.86, 9.18 and 12.45 at 25°C
NIST10
NIST10:1.68, 4.01, 6.86, 10.01 and 12.45 at 25°C
CUSTOM
CUSTOM: Use arbitrarily calibration value.
CAL Alarm
0 (OFF)
- 400 days Set the interval of calibration in days.
Temp. Type
ATC
ATC (Automatic Temperature Compensation):
The instrument detects the solution temperature
with the connected temperature sensor.
MTC
MTC (Manual Temperature Compensation): The
instrument performs temperature compensation
using the entered temperature.
Temperature
(Only MTC
Setup)
0.0 - 100.0°C
Default:
25.0
°C
Use arbitrarily temperature value when the temp.
Type MTC setting is active.
CAL Data Clear
Erase the calibration data.
Cartridge Info.
Use the Cartridge Information to update meter
settings for the model and SN of the pH sensor
cartridge.
Model
Enter a model name using the alphanumeric entry
screen (up to 10 characters)
SN
Enter a serial number using the alphanumeric
entry screen (up to 10 characters)
General Setup
refer to General
Setup on page 33
Use General Setup to customize settings for the
meter.
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30
Содержание Oakton 300 Series
Страница 1: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 4: ...Regulations III www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 11: ...Contents X www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 24: ...Viewing Calibration data 13 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Страница 62: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...