![enviromental express ABF5000 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/enviromental-express/abf5000/abf5000_operation-and-instruction-manual_2416053017.webp)
+1.800.343.5319 or +1.843.881.6560 envexp.com
Environmental Express
• 15
AutoBlock® Fill:
Operation and Instruction Manual
EPA Method 200.2, Revision 2.8
4.0 Procedure, Solids Sample Preparation
4.1 Sieve a dried sample using a 5-mesh polypropylene sieve and grind in a
mortar and pestle. Weigh a representative sample of 0.5 ±0.01g into a
digestion vessel.
4.2 Add 2mL of (1+1) HNO
3
and 5mL of (1+4) HCl. Cover with a ribbed watch
glass (SC505) or reflux cap (SC506),and place in the HotBlock at 95°C.
4.3 Heat sample for 30 minutes.
4.4 (Optional) Although step 3 is the final heating step for EPA Method 200.2,
if the sample is suspected of having a high concentration of organic
compounds, it is recommend to complete this step:
Add 2mL of 30% H
2
O
2
to the well-cooled sample. Allow the exothermic
reaction to occur (approximately 10 minutes) and place the sample back
in the HotBlock at a temperature of 10° less than the original set point for
an additional 30 minutes. The reaction with the H
2
O
2
raises the sample
temperature. Boiling should not occur if the temperature of the HotBlock
is lowered.
Note: H
2
O
2
helps aid in the breakdown of high organic compounds in the sample thus
creating a more complete digestion.
4.5 Using the polycarbonate transfer racks, remove samples from the HotBlock
and allow them to cool completely.
4.6 Bring sample volume to 50mL with DI water.
4.7 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)
to remove insoluble material.
Note: The filtration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs, stop filtration and allow sediments to
“settle out”. Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the filter into the digestate.
All QC samples, concentration limitations, elemental lists and reagent specifications
are addressed in depth in EPA Method 200.2. Safety concerns are also part of the full
method. Follow the instructions listed in EPA Method 200.2. These steps should only
be used as a guide to help improve the performance of your HotBlock.