Adjusting of the switch response time potentiometer P2

Adjust the switch response time with the potentiometer P2 according to your requirement between the range 
of 0.25 and 5 seconds. By turning P2 to left stop, the delay will be 0.25 seconds. At right stop the delay time 
will be 5 seconds. Thereby a fluttering of the relay can be stopped, which have been produced by critical 
adjustment of the sensitivity.

Measuring materials with low bulk density or low DK value

In order to detect threshold levels reliably even for materials with low bulk density or low DK values, the 
transmitter and receiver unit can also be installed on one side of the pipe (see chapter 4.3, Fig. 2).
In this installation variant, the opposite metallic wall is used as a reflection surface for the microwave. In 
this case the orientation of transmitter and receiver must not be exactly parallel to each other. Otherwise 
there is a risk that the emitted microwave will not be reflected to the receiver. A slightly angled positioning is 
required. Depending on “angle of dip = emergent angle”, a slightly angled positioning is required depending 
on the distance of the reflector wall. The distance travelled by the microwave is approximately twice as long 
with this arrangement compare to an normal installation. This mounting option can also be use in case of 
less space on one side of the container. The settings of the measuring point are then made as described in 
chapter 6.

7.  Fault clearance

Error

Cause

Action

No reaction on relay contact
when level reached

Low density of material

Measuring point set to maximum 

sensitive:
- P1 right stop
- P2 left stop
- S2 OFF

Check installation situation

Low DK-value of material

Transmitter and receiver doesn‘t 

see each other

Red LED 1 doesn‘t light up

No power supply

Check power supply (24 V DC)

Check installation situation

Transmitter and receiver doesn‘t 

see each other

8. Notice

 

A minimum distance of 15 cm to adjacent walls is recommended to avoid disturbing reflections.

 

In case of less installation space or material with low bulk density or low DK-value, installation on one 
side of the container is possible.

 

Various process adapters are available to protect the sensor, observe ATEX zones or solve difficult 
process conditions. 

Содержание ProGap 2.0

Страница 1: ...OPERATING INSTRUCTIONS ProGap 2 0 LEVEL DETECTION ENVEA Process GmbH Gutedelstraße 31 79418 Schliengen GERMANY Tel 49 0 7635 827248 0 info process envea global www envea global ...

Страница 2: ...Technical statement 4 4 Mounting and installation 5 4 1 A measuring point consists of the following components 5 4 2 Required equipment 5 4 3 Sensor installation 5 5 Electrical connection 6 5 1 Electrical connection receiver 6 5 2 Electrical connection transmitter 6 6 Start up procedure 7 7 Fault clearance 8 8 Notice 8 9 Technical data 9 ...

Страница 3: ...ave measure through non conductive materials such as plastics glass etc This makes it possible to detect with the sensor from the outside or through a window In this way the measurement can be completely decoupled from the process For example to measure aggressive abrasive or lumpy materials Various process adapters can be used to solve applications with temperatures up to 1000 C pressure up to 20...

Страница 4: ... not performed correctly may result in death or injury Attention This symbol is used in the operating manual to denote actions which may result in danger to property 3 3 Operational safety The measuring system may only be installed by trained authorised personnel During all maintenance cleaning and inspection work on the pipelines or ProGap 2 0 components make sure that the system is in an unpress...

Страница 5: ...same level It is imperative that the sensor socket is free of dirt and deposits If the measuring point is not directly installed the sensor sockets can be closed with a cap with G 1 thread It is important to ensure that the transmitter and receiver are given the same orientation and are aligned precisely to each other Polarization sticker After successful alignment transmitter and receiver unit mu...

Страница 6: ...ly be placed on the connector in the transmitter unit is additionally fastened with the existing screws This will ensure that the contacts will not disconnect even on vibrated places The electrical connection of the transmitter receiver system is as shown in Figures 3 and 4 L H 0 25s 5s P1 P2 LED1 LED2 S1 S2 5 4 3 2 1 Fig 4 Electrical connection transmitter 2 1 Fig 3 Electrical connection receiver...

Страница 7: ...ement of smaller distances and low density materials could be possible Potentiometer P1 Adjustment of the threshold based on the sensitivity Left stop lowest sensitive Right stop highest sensitivity P2 Setting for time delay on the relay contact Left stop 0 25 s Right stop 5 s Adjustment of the signal threshold with potentiometer P1 With the potentiometer P1 the signal threshold can be adjusted so...

Страница 8: ...pending on angle of dip emergent angle a slightly angled positioning is required depending on the distance of the reflector wall The distance travelled by the microwave is approximately twice as long with this arrangement compare to an normal installation This mounting option can also be use in case of less space on one side of the container The settings of the measuring point are then made as des...

Страница 9: ...tion approx 20 VA Current consumption approx 850 mA Relay contact Max switching capacity 250 V AC Max start up current 6 A Max breaking capacity 230 V AC 250 VA Max switching current DC1 3 110 220 V 3 0 35 0 2 A Min breaking capacity 500 mW 10 V 5 mA Response time 0 25 s 5 s continuously adjustable Measuring frequency K Band 24 125 GHz 100 MHz Transmitting power max 5 mW Weight Transmitter 1 1 kg ...

Отзывы: