Specification Information
Transportation and
storage conditions
▪
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪
Relative humidity range of 10% to 95%, non-condensing.
▪
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions
▪
The warmer must acclimate to the room temperature it will be placed—24 hours
is recommended.
▪
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪
The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
▪
Medical equipment classified by Underwriters Laboratories with respect
to electrical shock, fire, and mechanical hazards only, in accordance with
UL 60601-1 and CAN/CSA C22.2 No. 601.1.
▪
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
▪
IP-XO ordinary
Model
EC770L
Weight
Net: 173 lb (78 kg)
Ship: 258 lb (117 kg)
Storage cavity
capacity
72 1-liter bags, 24 bags per basket
54 1-liter bottles, 18 1-liter bottles per basket
Temperature range
Irrigation fluids (IRR): 98°F to 150°F (37°C to 66°C)
Injection fluids (INJ): 98°F to 104°F (37°C to 40°C)
Clearance
requirements
3” (76mm) from rear
1” (25mm) from top
1” (25mm) from sides
E471516
Protective Earth
Ground Symbol
Содержание EC770L
Страница 1: ...Operator s Manual Fluid Warmer EC770L MN 39935 REV 01 3 19 EN ...
Страница 4: ...FOREWORD Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 4 This page intentionally left blank ...
Страница 10: ...SAFETY Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 10 This page intentionally left blank ...
Страница 11: ...LABELS Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 11 ABELS Label Locations EC PHD 006516 a c d b ...
Страница 18: ...INSTALLATION Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 18 This page intentionally left blank ...
Страница 26: ...OPERATION Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 26 This page intentionally left blank ...
Страница 30: ...MAINTENANCE Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 30 This page intentionally left blank ...
Страница 38: ...TROUBLESHOOTING Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 38 This page intentionally left blank ...
Страница 41: ......
Страница 42: ......
Страница 43: ......