M
AINTENANCE
Fluid Warmer
▪
Operator’s Manual
▪
MN-39935
▪
Rev 1
▪
3/19
27
AINTENANCE
Maintenance Schedule
Daily
For daily maintenance, do the following.
Check:
▪
the air vents in the airflow insert panels are not obstructed (if applicable).
▪
all fan guards are clear and not obstructed (if applicable).
▪
the number of bottles, bags, or blankets do not exceed the maximum capacity
per shelf or basket.
Clean:
▪
any spills with a clean, lint free cloth.
Monthly
For monthly maintenance, do the following.
Check:
▪
the door gasket for tears and holes. Make sure that it is firmly attached to the
door. Check the seal when the door is closed.
▪
the guards around the air temperature sensor are in place and fully secure to the
warmer.
▪
the hardware securing the warmer(s) to the wall, if applicable.
Clean:
▪
the outside of the warmer with stainless steel cleaner.
▪
vacuum (if applicable):
□
fan openings
□
fan sail switch
□
vent openings
Yearly
For yearly maintenance, do the following.
Check:
▪
the set-point temperature compared to the actual temperature displayed.
□
Check the cavity air temperature with a thermocouple placed 1” (25mm) from
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
▪
the condition of the plug and cord and replace if damaged.
▪
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
▪
the shelves and interior of the warmer.
Содержание EC770L
Страница 1: ...Operator s Manual Fluid Warmer EC770L MN 39935 REV 01 3 19 EN ...
Страница 4: ...FOREWORD Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 4 This page intentionally left blank ...
Страница 10: ...SAFETY Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 10 This page intentionally left blank ...
Страница 11: ...LABELS Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 11 ABELS Label Locations EC PHD 006516 a c d b ...
Страница 18: ...INSTALLATION Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 18 This page intentionally left blank ...
Страница 26: ...OPERATION Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 26 This page intentionally left blank ...
Страница 30: ...MAINTENANCE Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 30 This page intentionally left blank ...
Страница 38: ...TROUBLESHOOTING Fluid Warmer Operator s Manual MN 39935 Rev 1 3 19 38 This page intentionally left blank ...
Страница 41: ......
Страница 42: ......
Страница 43: ......