T
ROUBLESHOOTING
Blanket Warmer
▪
Operator’s Manual
▪
M N-39924
▪
Rev 3
▪
3/20
46
E-11
ES11
ES21
ES31
ES41
ES51
ES61
ES71
Cavity sensor
Pad sensor 1
Pad sensor 2
Pad sensor 3
Pad sensor 4
Pad sensor 5
Pad sensor 6
Pad sensor 7
Sensor is open. Software disengages heating pads. Acknowledge error by
pressing the standby button. If error persists, a qualified service technician
should test the sensor.
◾
Test the sensor. Disconnect the sensor from the warmer control PCB
and use an Ohm meter to measure the resistance of the sensor.
Sensor PR-37140 should measure 10 KOhm ± 1.5 KOhm at 25°C (77°F).
Sensor SN-33541 should measure 100 Ohm ± 10 Ohm at 25°C (77°F).
Replace sensor if measurement is outside of tolerance.
◾
Check the wires for integrity. Check for proper and secure connections
at the control and terminal block. If necessary, re-secure the faulty
connections.
◾
Contact service if error persists.
E-30
Under
temperature
The blanket cavity temperature is lower than the set temperature for 90
minutes or longer.
◾
Make sure the door is closed.
◾
If the cavity is overloaded, redistribute the inventory. Do not exceed the
height of the insert.
◾
Test the sensor. Disconnect the sensor from the warmer control PCB
and use an Ohm meter to measure the resistance of the sensor.
Sensor PR-37140 should measure 10 KOhm ± 1.5 KOhm at 25°C (77°F).
Sensor SN-33541 should measure 100 Ohm ± 10 Ohm at 25°C (77°F).
Replace sensor if measurement is outside of tolerance.
*E-31
Cavity sensor
◾
The sensor reading is above the temperature set-point. Blanket warmer
triggers at 15° over set-point. Fluid warmer triggers at 5° over set-point.
◾
The difference between the room temperature and the fluid set-point
temperature must be greater than 11°C (20°F).
◾
Contact service if error persists.
P131
P231
P331
P431
P531
P631
P731
Pad sensor 1
Pad sensor 2
Pad sensor 3
Pad sensor 4
Pad sensor 5
Pad sensor 6
Pad sensor 7
Heater pad over-temp error.
◾
Software disengages heating pads.
◾
Acknowledge error by pressing the standby button.
◾
Allow the warmer to cool.
◾
Contact service if the error persists.
*E-33
Cavity sensor
◾
Sensor reading is above maximum allowable temperature set-point and
over temp value. Blanket warmers trigger at 82°C (180°F). Fluid warmers
trigger at 71°C (160°F).
◾
Contact service if error persists.
*E-50
Analog to Digital
Converter Error
◾
Remove inventory, discard if necessary, and allow the warmer to cool
down.
◾
If error persists after cool down and reset, the control assembly should
be replaced by a qualified service technician. Contact service.
Code
Refers to
Action required
Содержание Comfort EC250
Страница 4: ...FOREWORD Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 4 This page intentionally left blank...
Страница 6: ...FOREWORD Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 6 This page intentionally left blank...
Страница 10: ...SAFETY Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 10 This page intentionally left blank...
Страница 11: ...LABELS Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 11 L ABELS Label Locations EC PHD 004557 a c d b...
Страница 26: ...SPECIFICATIONS Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 26 This page intentionally left blank...
Страница 30: ...INSTALLATION Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 30 This page intentionally left blank...
Страница 42: ...MAINTENANCE Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 42 This page intentionally left blank...
Страница 50: ...TROUBLESHOOTING Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 50 This page intentionally left blank...
Страница 57: ......
Страница 58: ......
Страница 59: ......