M
AINTENANCE
Blanket Warmer
▪
Operator’s Manual
▪
M N-39924
▪
Rev 3
▪
3/20
38
Continued from previous page
Yearly
For yearly maintenance, do the following.
Check:
▪
the set-point temperature compared to the actual temperature displayed.
□
Check the cavity air temperature with a thermocouple placed 1” (25mm) from
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
▪
the condition of the plug and cord and replace if damaged.
▪
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
▪
the shelves and interior of the warmer. See topic
How to Clean the Warmer
for the
list of approved cleaners.
Содержание Comfort EC250
Страница 4: ...FOREWORD Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 4 This page intentionally left blank...
Страница 6: ...FOREWORD Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 6 This page intentionally left blank...
Страница 10: ...SAFETY Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 10 This page intentionally left blank...
Страница 11: ...LABELS Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 11 L ABELS Label Locations EC PHD 004557 a c d b...
Страница 26: ...SPECIFICATIONS Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 26 This page intentionally left blank...
Страница 30: ...INSTALLATION Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 30 This page intentionally left blank...
Страница 42: ...MAINTENANCE Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 42 This page intentionally left blank...
Страница 50: ...TROUBLESHOOTING Blanket Warmer Operator s Manual MN 39924 Rev 3 3 20 50 This page intentionally left blank...
Страница 57: ......
Страница 58: ......
Страница 59: ......