50455/RTEX/0807
UTImeter Rtex
39
9.4 Tape replacement
THE REPLACEMENT OF THE TAPE DOES NOT REQUIRE TO RE-CALIBRATE THE TEMPERATURE.
Follow the different sequences as described below. The Figure 12-1 : general assembly, list of the main
spare parts can also help.
9.4.1 Disconnecting the tape from the sensor
Follow the instructions of section 9.5 "Sensing
probe replacement".
9.4.2 Disconnecting
the
tape
from
the
electronic box
Unscrew with the 2.5 Allen key the 2 screws
(A) of the battery holder and pull it out as
shown on Figure 9-5.
Unscrew with the 2.5 Allen key the 4 screws
(B) of the display unit and pull it gently out as
shown on Figure 9-5.
Disconnect the connecting plug (C) as shown
on Figure 9-6 and remove the display unit.
Unscrew with the 2.5 Allen key the tape holder
(G) by removing the 2 screws (F) and the
grounding cable (D) as shown on Figure 9-6.
Do not loose the 2 remaining screws that
secure the reel axle.
(A)
(B)
Figure 9-5
(D) grounding cable of tape
(E) black oversleeve of tape
(G) tape holder
(H) tape
(C) connecting plug
to display unit
(I) securing screw
(F) securing screw
Figure 9-6
Содержание HERMetic UTImeter Rtex
Страница 55: ...50455 RTEX 0807 UTImeter Rtex 55 Figure 12 1 general assembly list of the main spare parts...
Страница 56: ...50455 RTEX 0807 UTImeter Rtex 56 Figure 12 2 display unit assembly TS 10210 details...
Страница 59: ...50455 RTEX 0807 UTImeter Rtex 59 Figure 12 5 storage tube SS1 Q1 TS 10184 details...
Страница 60: ...50455 RTEX 0807 UTImeter Rtex 60 Figure 12 6 storage tube SS1 Q2 TS 10183 details...
Страница 61: ...50455 RTEX 0807 UTImeter Rtex 61 Figure 12 7 storage tube SS2 Q2 with load TS 10182 details...
Страница 62: ...50455 RTEX 0807 UTImeter Rtex 62 Figure 12 8 tape cleaner details...
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......