59
13 Function Test
The Endopuls 811 will run a self-test that checks all internal components after it is switched on.
An error message is shown in case of faults.
In addition, a function test shall be made monthly or in case of doubt of the proper functioning of
the device.
NOTE:
Before performing the function test, check whether the hand-piece and the foot switch are
connected correctly to the device.
Check for proper mains connection.
Switch on the device.
Depress the footswitch briefly – the fan and shockwave generator will start immediately, whereby the
shockwave generator has to operate at the frequency indicated on the display (5 Hz as default value).
NOTE:
In conclusion of the test, switch off the device at the main switch.
If a treatment is to be performed immediately afterwards, set the required treatment parameters and
proceed as mentioned in Chapter 11.7
14 Error messages
Monitoring the hand-piece temperature (sww 12.1.2)
Generating mechanical shockwave energy causes a considerable build-up of
heat in the hand-piece. To avoid shortening the lifetime of the hand-piece, a
temperature switch has been integrated. This triggers an internal switch-off, if
the temperature becomes too high, forcing the hand-piece to cool.
If the temperature switch is activated, this is indicated by a message on the
display and shocks can no longer be emitted.
As soon as the hand-piece has reached the operating temperature,
automatically the pop-up disappears and the unit is ready to use again.
No SD-Card found
If the SD card is not inserted, the message ‘SD card not found’ appears when
the 'Favorites' and 'Memory' buttons are pressed.
Insert card and confirm with ‘OK’.