Enraf-Nonius B.V.
Curapuls 670 Service Manual
1403773
Maintenance
June 1997
4-9
Installation:
Installation of the front cover is in reverse order of the removal
procedure.
4.5.3
Top cover
Removal:
Remove the front cover (see par. 4.5.2). Remove the four screws at the
top at the inside of the housing. Disconnect the cables from CN1 and
CN2 from the main PC Board. Lift the top cover carefully, disconnect
the mains switch if necessary. Remove the top cover.
Installation:
Installation of the top cover is in reverse order of the removal procedure.
4.5.4
Main PC Board
Removal:
See par. 4.5.2 for removal of the front cover. Disconnect the coaxial
cables from output channels A (CN5) and B (CN4) and from connectors
CN1, CN2, CN3 CN8, CN11. Remove the 10 mounting screws.
Installation:
Installation of the main PC Board is in reverse order of the removal
procedure.
Caution:
Check that all fuses have the correct value (see par. 1.3.1)
Note:
Check when mounting the main PC Board for the right version software
(version 4 or above) of EPROM IC1, see appendix B, fig. 5.
Note:
When inserting the EPROM with the application software into the socket,
the PC Board must be supported to ensure that no components will be
damaged by mechanical force.
Note:
Check that jumper ST1 is removed.
4.5.5
User interface PC Board
Removal:
See par. 4.5.2 for removal of the front cover, and par. 4.5.3 for removal
of the top cover. Remove the mounting screws. Remove the user
interface / LCD display module- assembly. For removal of the LCD
display see par. 4.5.6.
Installation:
Installation of the user interface PC Board is in reverse order of the
removal procedure.
Note:
When inserting the EPROM with the control / language software into the
socket, the PC Board must be supported to ensure that no components
will be damaged by mechanical force.
4.5.6
LCD display module
Removal:
See par. 4.5.5 for removal of the user interface PC Board. Disconnect
Содержание 1403773
Страница 1: ...Curapuls 670 SERVICE MANUAL Part number 1403773 ...
Страница 2: ......
Страница 6: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Index June 1997 4 ...
Страница 16: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Test equipment and tools June 1997 3 2 ...
Страница 30: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Maintenance June 1997 4 14 ...
Страница 32: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Storage and transport June 1997 5 2 ...
Страница 38: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 2 Figure 2 Curapuls 670 top view ...
Страница 39: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 3 Figure 3 Curapuls 670 front view ...
Страница 48: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 12 ...
Страница 56: ......
Страница 58: ...No SI 670 02 Oct 1998 Page 2 of 2 Figure 1 Detail of the main PC Board ...
Страница 59: ......
Страница 60: ...Subsidiary of Henley Healthcare Inc Enraf Nonius B V P O Box 810 2600 AV Delft The Netherlands ...