Maintenance
June 1997
4-2
move a rectangle right to left between the two lines at the bottom of the LCD display
(otherwise see par. 4.4.1.7).
Entering the service mode:
Testing of other parts of the Curapuls 670 is done in the so called “SERVICE MODE”.
For starting the service mode enter the following key code when in control panel test
mode, press in sequential order the following keys:
2 x cursor up key,
1 x enter key,
1 x cursor down key,
2 x cursor right key.
After entering this code the Curapuls 670 is in service mode. The following options are
displayed:
Test number
:
test number of the test for the Curapuls 670.
Sub-selection :
test sub-selection if available.
Result hex.
:
test results in hex. if relevant.
Result dec.
:
test results in dec. if relevant.
Note:
Only the tests which are relevant for the functional test are shown below, other
tests are used during production, for diagnostic purposes or used for
adjustment/calibration (see par. 4.6).
Selection of a test number is done by rotating the central control. To start a test press
the start/stop key. When a test is executed the green LED above the start/stop key will
light. Stop a test by pressing the start/stop key again. During execution of a test the
test number can not be changed, first stop a test before changing to an other test.
Note:
Leaving the service mode is done by switching off the Curapuls 670.
4.3.3
Main board software version number
Enter the service mode and select test 36
To display the application software version press the start/stop key.
The displayed version is A4 or higher (A = application software Curapuls 670, 4 =
version number).
Stop the test by pressing the start/stop key.
Continue with the next test or leave the service mode by switching off the Curapuls 670.
4.3.4
User interface / language software version number
Enter the service mode and select test 53
The language software version is displayed at “Result hex” on the display.
The user interface software version is displayed a “Result dec” on the display.
Содержание 1403773
Страница 1: ...Curapuls 670 SERVICE MANUAL Part number 1403773 ...
Страница 2: ......
Страница 6: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Index June 1997 4 ...
Страница 16: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Test equipment and tools June 1997 3 2 ...
Страница 30: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Maintenance June 1997 4 14 ...
Страница 32: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Storage and transport June 1997 5 2 ...
Страница 38: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 2 Figure 2 Curapuls 670 top view ...
Страница 39: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 3 Figure 3 Curapuls 670 front view ...
Страница 48: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 12 ...
Страница 56: ......
Страница 58: ...No SI 670 02 Oct 1998 Page 2 of 2 Figure 1 Detail of the main PC Board ...
Страница 59: ......
Страница 60: ...Subsidiary of Henley Healthcare Inc Enraf Nonius B V P O Box 810 2600 AV Delft The Netherlands ...