INTENDED USE
WARNINGS AND CAUTIONS
Enova Illumination LED Surgical Headlight Systems are intended to be used by
surgeons, dentists, veterinarians, or other healthcare professionals as an adjunct
light source during clinical or surgical procedures in a medical setting. Enova LED
headlight output is measured at 14 inches (35cm) and powered using recharge-
able lithium-ion battery packs. The pivot bracket allows for the coaxial alignment of
the headlight. All Enova headlights and batteries are serialized. The serial number
for single ratchet headbands is located on the left side of the headband where
then comfort pad attaches to the headband. The serial number for double ratchet
headbands is located under the top ratchet comfort pad. The serial number for
the batteries is located on the back of the battery.
Before operation, Enova LED Surgical Headlight Systems users should read and
understand the operation manual for all the equipment used in conjunction with
the headlight. Contact Enova Illumination or your local distributor for service and
warranty information.
• DO NOT look directly into the headlight when illuminated. DO NOT shine light
directly into the patient's eyes or face. Due to the intensity of light, eye injury
may result. The patient’s eyes should be appropriately protected in surgeries
involving the face or head.
• After a few minutes of operation, the headlamp housing may become warm
to the touch. This is normal and functions to dissipate the heat from the LED.
Always use a surgical glove or sterile cloth to adjust the angle of the lens when
in use.
• Use only Enova supplied batteries and chargers. Using other chargers or bat-
teries may cause fire, explosion, or damage to the battery packs or headlights
and void the warranty. The battery pack must be powered off when charging.
DO NOT USE OR STORE A BATTERY CHARGER IN THE OPERATING
ROOM OR SURGERY SUITE.
• All Enova Headlights are wireless. Do not use the headlight while batteries
are charging.
• DO NOT use the headlight, batteries, or charger in explosive gas environ-
ments.
• DO NOT autoclave, immerse or submerge in liquid, spray with liquids or leave
the light on when in the carrying case as it could cause damage to the head-
light.
•
DO NOT INTERCHANGE BATTERY CABLES. Use specified cable for correct
headlight; otherwise, the headlight will function abnormally and damage the
circuitry causing the light to fail.
2