Preventative Maintenance Checklist 

 

Maintenance  of  the  table  is  responsibility  of  the  owner.    Manufacturer  recommends 
following areas be inspected by certified technician every six months to assure proper 
operation  of  table.  Copy  of  this  checklist  should  be  used  for  each  inspection  and 
maintained in table’s master file.  Each inspection area is required to indicate a pass or 
fail  status.    ALL  AREAS  failing  inspection  must  be  serviced/repaired  immediately  by 
certified technician.  See ENOCHS Service Manual for Power Encore Series Tables for 
repair/service  advice  and  instructions.  Direct  all  questions  to  ENOCHS  Technical 
Support at 1-800-428-2305.  

 

 

Serial Number:

  

  

EL 

Inspection 

Date: 

 

 

PASS  

 

FAIL 

 

INSPECTION AREA  

 

 

grounding continuity 

 

 

mains power cord and electrical plug free from all nicks, cuts, abrasions or 
any other damage. 

 

 

mains power cord ON/OFF switch operates. 

 

 

mains power indicator light on front face plate functional. 

 

 

t

able’s receptacles functional (applicable on 4550 & 4520 with this option) 

 

 

Instrument warming drawer indicator ON/OFF switch functional.                        
(applicable on 4550 & 4220 with warming drawer option) 

 

 

warming drawer functional (applicable on 4550 & 4520 with this option) 

 

 

all foot control pedals move freely, returning to neutral position  without 
sticking (applicable on all tables with this option) 

 

 

foot control power cord free of damage and firmly held in place by strain 
relief at back of foot control and at table base 

 

 

TABLE  UP/DN functions exhibit smooth range of movement 

 

 

BACK  UP/DN functions exhibit smooth range of movement 

 

 

pelvic tilt adjusts smoothly and easily 

 

 

leg extension moves freely in and out of table 

 

 

irrigation basin drawer moves freely in and out of table 

 

 

stirrup length and lateral adjustments have smooth range of movement 

 

 

upholstery is free from cuts, rips or punctures 

 

 

accessories  rails (if applicable) are secure 

 

 

accessories  in good working order  

 

PRODUCT AT A GLANCE 

19 

Range of Motion

 

 
Power Back: 0º horizontal to +80º   =± 2º  
Power Height: Min:  

18.5” ± .25 in.   Max: 38” ± .25 in.

  

 

Stainless Steel 
Irrigation basin  

Perineal 

recess 

Adjustable stirrups   

Pelvic tilt 

Range of Motion

 

Min:  Flat  

      Max.: 11º ± 2º

 

Pull-out 

leg extension

 

Paper cutter 

Paper roll holder  

Instrument warmer 
drawer 

(optional on models 

4550 & 4520 only) 

Power back   
 

Side drawer 

storage 

(not 

available on 4510)

 

 

Stainless steel drawer  

(standard on all models)  

Mains Power light 

Mains Power Input Module 

Instrument Warming 

Drawer ON/OFF SWITCH 

(on models  with optional 

warmer drawer )

 

Power Height  

 

Foot control 

(standard on models 4520 & 4510,                           

optional on model 4550) 

Содержание Power 4510

Страница 1: ...nance 17 contact information 3 log 22 semi annual checklist 19 questions 3 Malfunction of table 5 Operational conditions 4 Paper cutter 14 Paper roll holder 14 Patient load 8 Pelvic tilt 14 adjustment 14 Power features 9 warnings 9 11 Perineal recess 14 Product at a glance 6 7 Purchase record 23 Range of motion 6 Safety information 5 Symbol definitions 4 Setting up table 8 Specifications 21 Stirru...

Страница 2: ...Vacuum Formed Top 16 Table Maintenance and Cleaning Recommendations 17 Upholstery Care 18 Preventative Maintenance Checklist 19 Troubleshooting 20 Product Specifications 21 Table Maintenance History 22 Guarantee of Quality Warranty Index 23 INDEX 24 ENOCHS Guarantee of Quality ENOCHS Inc guarantees ENOCHS products to be free from defects in material or workmanship under normal installation use and...

Страница 3: ... professionals Tables are designed to position and support patients during procedures DISPOSAL OF EQUIPMENT At the end of the product life consult local codes and ordinances for proper disposal of table accessories and other consumable goods Note products may have become contaminated from normal use MAINTENANCE AND SERVICE QUESTIONS Contact ENOCHS with questions concerning this manual proper opera...

Страница 4: ...ual electrical receptacles with circuit breaker and reset option for 4550 and 4520 only Instrument warmer located in foot end drawer option for 4550 and 4520 only Instrument warmer ON OFF switch located on front side of table illuminates when warming unit is ON Switch has internal circuit breaker with reset Irrigation basin 2 5 deep stainless steel basin Mains power entry module IEC connection int...

Страница 5: ...f tripped reset and test table functions no additional action required If table does not function call for service The Encore Power Series Tables are designed for intermittent operation with a duty cycle of 1 minute ON 9 minutes OFF DO NOT EXCEED duty cycle doing so could damage system WARNING To avoid a fire hazard or risk of electric shock this table must only be connected to a supply mains with...

Страница 6: ...h this option all foot control pedals move freely returning to neutral position without sticking applicable on all tables with this option foot control power cord free of damage and firmly held in place by strain relief at back of foot control and at table base TABLE UP DN functions exhibit smooth range of movement BACK UP DN functions exhibit smooth range of movement pelvic tilt adjusts smoothly ...

Страница 7: ...ld water rinse THEN Wipe entire surface with cloth damp with bleach solution followed by a clear cold water rinse Surface Mildew 1 Tablespoon bleach to one quart water Sponge area with solution until mildew is removed Follow with a clear cold water rinse Ball point Ink Isopropyl alcohol 70 followed immediately by several clear water rinses IMMEDIATELY remove mark using full strength Isopropyl alco...

Страница 8: ...emi annual Maintenance Inspection Checklist page 19 CLEANING RECOMMENDATIONS S PE CIAL CL EA NI NG A LE RT Follow guidelines for cleaning procedures set forth by the CDC reference Isolation Precautions OSHA reference Bloodborne Pathogen Standards APIC Association for Practitioners in Infection Control and any other appropriate federal or state agencies PAINTED OR STAINLESS STEEL SURFACES NEVER USE...

Страница 9: ...through opening and secure laminate to table 4 Insert drawer into table VACUUM FORMED TOP Top is 100 seamless and manufactured without crevices or decorative embossing that collect dirt and makes cleaning difficult Top is padded with high density foam with additional contoured padding on the sides for patient comfort Top is constructed with a steel frame and backing designed for strength and durab...

Страница 10: ... desired length To Store Push unit gently back under seat section of table Irrigation Basin Access push extension into stored position lifting slightly while pushing under seat section for direct access to basin STIRRUPS Concealed when not in use the multi position stirrups offer length as well as lateral adjustment to accommodate each patient Stirrup locks into position with weight of patient s f...

Страница 11: ...4550 s Programmable Hand Control has three memory positions along with independent controls to adjust table s HEIGHT from 18 5 to 38 and the couch BACK to various degrees from horizontal to an 80º chair position To install Connect hand control cable to input port on the back face plate of table and secure with fasteners DO NOT CONNECT ANY OTHER HAND CONTROL INTO TABLE S INPUT PORT To disconnect Un...

Страница 12: ...ns power connection at the base of the table provides power for table height and back adjustment on all models It also provides power to the dual electrical receptacles and instrument warmer available on the 4550 4520 The module consists of an ON OFF toggle switch internal circuit breaker and IEC input Connect Mains Power Plug the power cord s IEC end into the Mains Power Entry Module then plug co...

Отзывы: