Back adjustment 

 

Raise  Back:

    Place  foot  firmly  on  center  of  pedal 

marked 

BACK

.    Depress  left  side  of  pedal 

designated  by 

 

UP

.    When  desired  angle  is 

reached, release pressure from pedal.   

 
Lower  Back:

  Place  foot  firmly  on  center  of  pedal

 

marked 

BACK

.    Depress  right  side  of  pedal 

designated  by 

DN   

.  When  desired  angle  is 

reached, release pressure from pedal.   

 

 

STANDARD HAND CONTROL 

 

The  ENCORE  4545  Hand  Control  is  an  option  for  model 
4520.    Hand  Control 

allows  fast  easy  adjustment  of  table’s 

HEIGHT  from  18.5

”  to  38"  and  the  couch  BACK  to  various 

degrees from horizontal to an 80º chair position.  

 

DO  NOT  CONNECT  ANY  OTHER HAND  CONTROL  INTO 
INPUT PORT. 

To  install:    Connect  hand  control  cable  to  input  port  on  the 
back face plate of table and secure with fasteners.  
To  disconnect:  Unscrew  fasteners,  securely  grasp  plug  and 
pull  from  port. 

DO  NOT  PULL 

plug  out  with

 

hand  control

 

cable

.  

 

 

 

 
 
 
STANDARD HAND CONTROL OPERATION 

 

 

                                           

Height Adjustment

  

 

Raise  Table

:  Press  button  marked 

TABLE       

When  desired  height  is  reached,  release 
pressure from button. 

 

Lower  Table:

  Press  button  marked 

TABLE     

When  desired  height  is  reached,  release 
pressure from button

   

                                           

Back Adjustment 

Raise  Back:

    Press  button  marked 

BACK     

When  desired  angle  is  reached,  release 
pressure from button.   

 
Lower Back:

 Press button marked BACK

     

.  

When desired angle is reached, release pressure 
from button

 

 

 

 

 

 

PULL-OUT LEG EXTENSION 

The extra wide padded leg extension moves in and out 
from under the seat section. Extension is padded and 
upholstered in durable vinyl.  Independent operation 
allows fast transition between leg extension and basin. 
Patients with limited hip rotation may be positioned with 
feet on extension while still allowing full access to basin 
by practitioner.

 

To Extend:

 Grasp front edge of extension and pull out to desired length. 

To Store: 

Push unit gently back under seat section of table. 

Irrigation  Basin  Access: 

push  extension  into  stored  position  lifting  slightly 

while pushing under seat section for direct access to basin. 

 

 

 

 
 

STIRRUPS

 

 

Concealed when not in use, the multi-position stirrups offer 
length  as  well  as  lateral  adjustment  to  accommodate  each 
patient.  Stirrup  locks  into  position  wi

th  weight  of  patient’s 

foot.  Stirrup  shoes  are  coated  in  durable  Plastisol  for 
efficient cleaning and patient comfort 

 

 
 
 
 

 
 

STIRRUP ADJUSTMENT 

 

Lateral Adjustment: 

Grasp stirrup bars, located under seat section and pull out to 

full bar extension.  Unfold stirrup shoes.  Lift bar approximately ½"(1.3cm) and 
move outward, lowering to desired angle.  

Length Adjustment:

  First make desired lateral adjustment of stirrups.  

To  Decrease  Stirrup  Bar  Length

:    Firmly  push  stirrup  bar  into  table  stopping  at 

desired length. Stirrup bar automatically locks, prohibiting increase in length.   

To  Increase  Stirrup  Bar  Length

:  Remove  weight  from  stirrups,  lift  bar 

approximately ½ " (1.3cm), return to center position and bring to full extension. 
Make lateral adjustment to desired position then push bar into table to desired 
length.

 

Perineal Angle: 

To increase patient’s leg angle for great perineal access, decrease 

stirrup bar length creating a shorter distance from stirrup shoe to end of table.  

Stirrup  Storage:

  Lift  bars  approximately  ½"  (1.3cm)  and  move  to  center  position.      

Fold down stirrup shoes against bar and gently slide bars under seat section. 

 

 

10 

15 

 

 

WARNING:

  Before activating hand control, confirm area around 

all moving parts of the table is clear of any obstructions to avoid 
product damage and/or bodily injury.

 

 

 

 

 

CAUTION:

 Confirm stirrup bars are secure in 

the lateral position before using. 
 
Stirrups  are  designed  to  support  patient’s 
legs, NOT 

entire patient’s weight

 

 

 

 

 

WARNING:

  Do not use leg extension to support entire patient weight.  

Extension designed to support weight of lower portion of patient’s legs.

 

 

 

 

Содержание Power 4510

Страница 1: ...nance 17 contact information 3 log 22 semi annual checklist 19 questions 3 Malfunction of table 5 Operational conditions 4 Paper cutter 14 Paper roll holder 14 Patient load 8 Pelvic tilt 14 adjustment 14 Power features 9 warnings 9 11 Perineal recess 14 Product at a glance 6 7 Purchase record 23 Range of motion 6 Safety information 5 Symbol definitions 4 Setting up table 8 Specifications 21 Stirru...

Страница 2: ...Vacuum Formed Top 16 Table Maintenance and Cleaning Recommendations 17 Upholstery Care 18 Preventative Maintenance Checklist 19 Troubleshooting 20 Product Specifications 21 Table Maintenance History 22 Guarantee of Quality Warranty Index 23 INDEX 24 ENOCHS Guarantee of Quality ENOCHS Inc guarantees ENOCHS products to be free from defects in material or workmanship under normal installation use and...

Страница 3: ... professionals Tables are designed to position and support patients during procedures DISPOSAL OF EQUIPMENT At the end of the product life consult local codes and ordinances for proper disposal of table accessories and other consumable goods Note products may have become contaminated from normal use MAINTENANCE AND SERVICE QUESTIONS Contact ENOCHS with questions concerning this manual proper opera...

Страница 4: ...ual electrical receptacles with circuit breaker and reset option for 4550 and 4520 only Instrument warmer located in foot end drawer option for 4550 and 4520 only Instrument warmer ON OFF switch located on front side of table illuminates when warming unit is ON Switch has internal circuit breaker with reset Irrigation basin 2 5 deep stainless steel basin Mains power entry module IEC connection int...

Страница 5: ...f tripped reset and test table functions no additional action required If table does not function call for service The Encore Power Series Tables are designed for intermittent operation with a duty cycle of 1 minute ON 9 minutes OFF DO NOT EXCEED duty cycle doing so could damage system WARNING To avoid a fire hazard or risk of electric shock this table must only be connected to a supply mains with...

Страница 6: ...h this option all foot control pedals move freely returning to neutral position without sticking applicable on all tables with this option foot control power cord free of damage and firmly held in place by strain relief at back of foot control and at table base TABLE UP DN functions exhibit smooth range of movement BACK UP DN functions exhibit smooth range of movement pelvic tilt adjusts smoothly ...

Страница 7: ...ld water rinse THEN Wipe entire surface with cloth damp with bleach solution followed by a clear cold water rinse Surface Mildew 1 Tablespoon bleach to one quart water Sponge area with solution until mildew is removed Follow with a clear cold water rinse Ball point Ink Isopropyl alcohol 70 followed immediately by several clear water rinses IMMEDIATELY remove mark using full strength Isopropyl alco...

Страница 8: ...emi annual Maintenance Inspection Checklist page 19 CLEANING RECOMMENDATIONS S PE CIAL CL EA NI NG A LE RT Follow guidelines for cleaning procedures set forth by the CDC reference Isolation Precautions OSHA reference Bloodborne Pathogen Standards APIC Association for Practitioners in Infection Control and any other appropriate federal or state agencies PAINTED OR STAINLESS STEEL SURFACES NEVER USE...

Страница 9: ...through opening and secure laminate to table 4 Insert drawer into table VACUUM FORMED TOP Top is 100 seamless and manufactured without crevices or decorative embossing that collect dirt and makes cleaning difficult Top is padded with high density foam with additional contoured padding on the sides for patient comfort Top is constructed with a steel frame and backing designed for strength and durab...

Страница 10: ... desired length To Store Push unit gently back under seat section of table Irrigation Basin Access push extension into stored position lifting slightly while pushing under seat section for direct access to basin STIRRUPS Concealed when not in use the multi position stirrups offer length as well as lateral adjustment to accommodate each patient Stirrup locks into position with weight of patient s f...

Страница 11: ...4550 s Programmable Hand Control has three memory positions along with independent controls to adjust table s HEIGHT from 18 5 to 38 and the couch BACK to various degrees from horizontal to an 80º chair position To install Connect hand control cable to input port on the back face plate of table and secure with fasteners DO NOT CONNECT ANY OTHER HAND CONTROL INTO TABLE S INPUT PORT To disconnect Un...

Страница 12: ...ns power connection at the base of the table provides power for table height and back adjustment on all models It also provides power to the dual electrical receptacles and instrument warmer available on the 4550 4520 The module consists of an ON OFF toggle switch internal circuit breaker and IEC input Connect Mains Power Plug the power cord s IEC end into the Mains Power Entry Module then plug co...

Отзывы: