This indicates that the depth is 74.45 feet, the probe temp is 76.4 F, the battery voltage
is 5.97, and the error code is 2 meaning that the battery voltage is below 6.5 volts.
SYSTEM STATE - A "?" command sent to the unit instructs the unit to send its current
operating state. This will include 3 lines of data. The first line reports the model number
and features, the software version, the unit serial number, and the user set ID number.
The second and third lines report the applicable system settings prefixed by the
commands used to set them as listed in the table in the section on external commands.
Example:
WW660 > ?
Mod #WW710-077303 Ver # 7101-008 Ser #080628015 ID #001
rn 25.0 rx 500.0 w 60.5 u0 a0 al 500.00 ah 25.00
nh 25.00 nl 500.00 cs0681n cm0681n RTU 11 cf0 6
WW710 >
In this example, "rn" indicates that the min range is 25.0, and the well temperature is
set to 60.5 degrees, etc.
MODBUS COMMUNICATIONS
The RS485 port on the Well Watch is configured to communicate as a slave device on
a two wire Modbus bus. When the RS485 Modbus port is enabled, the Well Watch
becomes a Modbus slave device listening for data requests from a Modbus master
device using either RTU or ASCII mode. Data requests with the address set in the Well
Watch will be processed and a data packet containing the requested data will be sent
back to the master.
The Well Watch can be configured for a Modbus network by setting the Serial mode to
RS485. Then select the Modbus mode, either RTU or ASCII. Then set the serial
settings, the baud rate, number of data bits, stop bits and the parity. Scroll the setting
once more to set the MODBUS address for the Well Watch. This can be any number
from 1 to 243 which has not been used by any other device on the network.